FDA Medical Device Recalls (openFDA)
Misonix, Inc: LySonix 2000 Ultrasonic Surgical Systems, MODEL NO. LY-2000. --- Old labeling: MANUFACTURED BY: MISONIX, INC. EXCLUSIVELY FOR SALE BY MENTOR CORPORATION UNDER EXCLUSIVE LICENSE FROM SONIQUE SURGICAL, INC. MISONIX, INC. 1938 NEW HIGHWAY, FARMINGDALE, NY 11735. Device was manufactured from September 1996 to present.
- Recall number
- Z-0070-04
- Recalling firm
- Misonix, Inc
- Product
- LySonix 2000 Ultrasonic Surgical Systems, MODEL NO. LY-2000. --- Old labeling: MANUFACTURED BY: MISONIX, INC. EXCLUSIVELY FOR SALE BY MENTOR CORPORATION UNDER EXCLUSIVE LICENSE FROM SONIQUE SURGICAL, INC. MISONIX, INC. 1938 NEW HIGHWAY, FARMINGDALE, NY 11735. Device was manufactured from September 1996 to present.
- Status
- Terminated
- Initiated
- 2003-08-13
- Posted
- 2003-10-30
- Terminated
- 2005-02-07
- Reason
- Firm received complaints of tingling or static shock sensation to end users caused by forcing the output connector into its receptacle misaligned. The misalignment causes additional voltage to appear on the patient probe.
- Root cause
- Other
- Action
- Misonix, Inc. notified the sole direct consignee, Byron Medical Inc., of the "Corrective Action" by letter on 8/13/03. On 9/17/03, Byron Medical Inc. notified the end users by letter, and provided instructions for proper assembly and warning labels to be affixed to the units.
- Quantity
- 1119 units (potentially)
- Distribution
- 1 distributor in Tucson, AZ, for further distribution to end users (physicians and medical facilities) nationwide & Puerto Rico, and in the following countries: Canada, Mexico, Australia, Switzerland, Israel, Saudi Arabia, Cyprus, Greece, Turkey, Germany, Dominican Republic, Sweden, Netherlands, China, Taiwan ROC, and S. Korea.
- Lot, serial or model codes
- Serial numbers between 0100 and 1219 (inclusive)
- 510(k) numbers
- ["K933559"]
- PMA numbers
- not on file
- Product code
- LFL
- Firm FEI
- 2435119
- Firm city
- Farmingdale
- Firm state
- NY
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28