Skip to the content

FDA Medical Device Recalls (openFDA)

Misonix, Inc: LySonix 2000 Ultrasonic Surgical Systems, MODEL NO. LY-2000. --- Old labeling: MANUFACTURED BY: MISONIX, INC. EXCLUSIVELY FOR SALE BY MENTOR CORPORATION UNDER EXCLUSIVE LICENSE FROM SONIQUE SURGICAL, INC. MISONIX, INC. 1938 NEW HIGHWAY, FARMINGDALE, NY 11735. Device was manufactured from September 1996 to present.

Recall number
Z-0070-04
Recalling firm
Misonix, Inc
Product
LySonix 2000 Ultrasonic Surgical Systems, MODEL NO. LY-2000. --- Old labeling: MANUFACTURED BY: MISONIX, INC. EXCLUSIVELY FOR SALE BY MENTOR CORPORATION UNDER EXCLUSIVE LICENSE FROM SONIQUE SURGICAL, INC. MISONIX, INC. 1938 NEW HIGHWAY, FARMINGDALE, NY 11735. Device was manufactured from September 1996 to present.
Status
Terminated
Initiated
2003-08-13
Posted
2003-10-30
Terminated
2005-02-07
Reason
Firm received complaints of tingling or static shock sensation to end users caused by forcing the output connector into its receptacle misaligned. The misalignment causes additional voltage to appear on the patient probe.
Root cause
Other
Action
Misonix, Inc. notified the sole direct consignee, Byron Medical Inc., of the "Corrective Action" by letter on 8/13/03. On 9/17/03, Byron Medical Inc. notified the end users by letter, and provided instructions for proper assembly and warning labels to be affixed to the units.
Quantity
1119 units (potentially)
Distribution
1 distributor in Tucson, AZ, for further distribution to end users (physicians and medical facilities) nationwide & Puerto Rico, and in the following countries: Canada, Mexico, Australia, Switzerland, Israel, Saudi Arabia, Cyprus, Greece, Turkey, Germany, Dominican Republic, Sweden, Netherlands, China, Taiwan ROC, and S. Korea.
Lot, serial or model codes
Serial numbers between 0100 and 1219 (inclusive)
510(k) numbers
["K933559"]
PMA numbers
not on file
Product code
LFL
Firm FEI
2435119
Firm city
Farmingdale
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register