Skip to the content

FDA Medical Device Recalls (openFDA)

ICU Medical, Inc.: ICU Medical 14" (36 cm) Appx 3.6 ml, Bifuse Transfer Set w/Rotating Luer, REF: 14237-28 The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein).

Recall number
Z-0069-2019
Recalling firm
ICU Medical, Inc.
Product
ICU Medical 14" (36 cm) Appx 3.6 ml, Bifuse Transfer Set w/Rotating Luer, REF: 14237-28 The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein).
Status
Completed
Initiated
2018-08-20
Posted
not on file
Terminated
not on file
Reason
Design change implemented changing the device from a blood set with a filter to a transfer set without a filter. Both devices had the same listing number. Customers may not be aware of the product design change and use the transfer set as a blood transfusion set without realizing a blood filter is not included.
Root cause
Device Design
Action
On 08/20/18, the firm, icumedical, mailed "URGENT MEDICAL DEVICE RECALL" l notices to consignees. The notice described the product, problem and actions to be taken. If product has been further distributed, the firm requested consignees to notify users. Customers were advised to return affected product only if the blood filter is needed. Customers needing response forms and/or return labels were advised to call 1-800-479-0551. For further inquiries call: 1-866-829-9025.
Quantity
2361 Units
Distribution
U.S. Distribution in states of: NY, AL, CA, MS, PA, GA, MI, NC, WI, IL, WA, UT and FL.
Lot, serial or model codes
lots: 85-540-SJ, 86-189-SL, 86-911-SJ
510(k) numbers
["K964435"]
PMA numbers
not on file
Product code
FMG
Firm FEI
2025816
Firm city
San Clemente
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register