Skip to the content

FDA Medical Device Recalls (openFDA)

Merge Healthcare, Inc.: eFilm Workstation 4.0 and 4.0.1. eFilm Workstation is one of the components of a PACS (Picture Archiving and Communications System). Product Usage: eFilm Workstation is a software application that is used for viewing medical images. eFilm Workstation receives digital images and data from various sources (including but not limited to CT,MR, US, RF units, computed and direct radiographic devices, secondary capture devices, scanners, imaging gateways or imaging sources). Images are stored, communicated, processed and displayed on the local disc of a workstation and/or across computer networks at distributed locations. Tasks that users may perform when viewing images include, but are not limited to: adjustment of window width and level; image stacking; annotation and measurement of regions of interest; and inversion, rotation, and flips of images. In addition, eFilm Workstation can be integrated with an institution s existing HIS or RIS for a fully integrated electronic patient record.

Recall number
Z-0068-2015
Recalling firm
Merge Healthcare, Inc.
Product
eFilm Workstation 4.0 and 4.0.1. eFilm Workstation is one of the components of a PACS (Picture Archiving and Communications System). Product Usage: eFilm Workstation is a software application that is used for viewing medical images. eFilm Workstation receives digital images and data from various sources (including but not limited to CT,MR, US, RF units, computed and direct radiographic devices, secondary capture devices, scanners, imaging gateways or imaging sources). Images are stored, communicated, processed and displayed on the local disc of a workstation and/or across computer networks at distributed locations. Tasks that users may perform when viewing images include, but are not limited to: adjustment of window width and level; image stacking; annotation and measurement of regions of interest; and inversion, rotation, and flips of images. In addition, eFilm Workstation can be integrated with an institution s existing HIS or RIS for a fully integrated electronic patient record.
Status
Terminated
Initiated
2014-08-28
Posted
2014-10-15
Terminated
2017-06-01
Reason
There is an issue related to eFilm Workstation versions 4.0 and 4.0.1 when having multiple studies open and utilizing the thumbnail panel to select multiple series from multiple studies may lead to the selection of an incorrect study.
Root cause
Device Design
Action
Merge sent an Important Product Information Notice letter, dated August 22, 2014 to consignees. The letter stated the issue, identified affected product, provided instructions, and told of a product update expected to be available by September 15, 2014. Customers with questions should contact Customer Service at 877-741-5369
Quantity
2,630
Distribution
Worldwide Distribution - US Nationwide and the countries of : Argentina, Australia, Canada, Chile, France, Germany, Hungary, Iceland, Italy, Lithuania, Mexico, New Zealand, Romania, Russia, South Africa, Sweden, and Yemen.
Lot, serial or model codes
versions 4.0 and 4.0.1
510(k) numbers
["K012211","K020995"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
1000116130
Firm city
Hartland
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register