Johnson & Johnson Surgical Vision Inc: OptiMedica Catalys Precision Laser System, CE 0044 containing software cOS 5.00.33 Product Label for Serial Numbers: 44109015; 44106215; 44107315; and 44065613 label states "Manufactured for: OptiMedica Corporation 1310 Moffett Park Drive Sunnyvale, CA 94089 USA Product Label for Serial Numbers: 46120916; 46123116, 46124617; 46131418, and 46131618 label states "Manufactured for: OptiMedica Corporation 510 Cottonwood Drive Milpitas, CA 95035 USA The Catalys Precision Laser System is indicated for use in patients undergoing cataract surgery for removal of the crystalline lens. Intended uses in cataract surgery include anterior capsulotomy, phacofragmentation, and the creation of single plane and multi-plane arc cuts/incisions in the cornea, each of which may be performed either individually or consecutively during the same procedure.

Recall number
Z-0066-2019
Recalling firm
Johnson & Johnson Surgical Vision Inc
Product
OptiMedica Catalys Precision Laser System, CE 0044 containing software cOS 5.00.33 Product Label for Serial Numbers: 44109015; 44106215; 44107315; and 44065613 label states "Manufactured for: OptiMedica Corporation 1310 Moffett Park Drive Sunnyvale, CA 94089 USA Product Label for Serial Numbers: 46120916; 46123116, 46124617; 46131418, and 46131618 label states "Manufactured for: OptiMedica Corporation 510 Cottonwood Drive Milpitas, CA 95035 USA The Catalys Precision Laser System is indicated for use in patients undergoing cataract surgery for removal of the crystalline lens. Intended uses in cataract surgery include anterior capsulotomy, phacofragmentation, and the creation of single plane and multi-plane arc cuts/incisions in the cornea, each of which may be performed either individually or consecutively during the same procedure.
Status
Terminated
Initiated
2018-05-09
Posted
not on file
Terminated
2021-01-13
Reason
Software upgrade exhibits failures when executing Daily Alignment Verification (DAV), specifically, cuts on the plastic hemisphere during DAV do not match the intended cuts shown on the treatment overlay screen on the system GUI.
Root cause
Device Design
Action
On May 9, 2018, affected customers in the U.S. and Puerto Rico were contacted by phone and verbally informed to stop using the affected system. On May 22, 2018, the firm, Johnson & Johnson Vision, sent an "URGENT MEDICAL DEVICE FIELD REMOVAL" letter dated May 18, 2018 via Federal Express overnight delivery with signature required to all customers with a system containing the affected software installed. The customers were instructed to: STOP Using the affected system and complete and return the Customer Reply Form via fax to: Johnson & Johnson Vision Quality Assurance at 1-714-247-4510, emailed to RegComplione@its.jnj.com, or provide to your Johnson & Johnson Vision Representative within 3 business days of receipt of the letter. The firm will coordinate the removal of the affected software for all affected systems. All impacted customers will be provided with the new/updated software version. If you have any questions, contact the Director, Quality Assurance at 408-273-4100 or email: creyes51@its.jnj.com.
Quantity
9
Distribution
US Distribution to states of: CO, GA, IL, MD including Puerto Rico and internationally to countries of: Austria and Germany.
Lot, serial or model codes
Model: Catalys Precision Laser System, CE 0044:  Catalys-U (US version) Serial Numbers: 44065613 - software cOS 5.00.33 install date 4/27/2018 46131618 - software cOS 5.00.33 install date 4/23/2018 46131418 - software cOS 5.00.33 install date 4/30/2018 44107315 - software cOS 5.00.33 install date 5/1/2018 46120916 - software cOS 5.00.33 install date 4/23/2018 46123116 - software cOS 5.00.33 install date 5/2/2018 44106215 - software cOS 5.00.33 install date 5/3/2018  Catalys-I (International version) Serial Number: 44109015 - software cOS 5.00.33 install date 5/1/2018 46124617 - software cOS 5.00.33 install date 5/3/2018
510(k) numbers
["K113479","K121091","K141079","K170322","K172002"]
PMA numbers
not on file
Product code
OOE
Firm FEI
3006695864
Firm city
Milpitas
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register