FDA Medical Device Recalls (openFDA)
GE Healthcare Beijing , Hangwei Medical Systems, Co. LTD: GE Brivo OEC715/785/865 C-Arm, Models: 715 Prime, 785 Essential, and 865 Plus. The QEC Brivo Mobile C-Arm X-Ray Products are designed to provide digital spot film imaging and fluoroscopic image guidance for all adult and pediatric populations for orientations between patient anatomy and surgical instruments. The product is used for general surgical applications and musculoskeletal procedures to visualize, for example, implant localizations or needle positions for aspirations,injections or biopsy.
- Recall number
- Z-0066-2014
- Recalling firm
- GE Healthcare Beijing , Hangwei Medical Systems, Co. LTD
- Product
- GE Brivo OEC715/785/865 C-Arm, Models: 715 Prime, 785 Essential, and 865 Plus. The QEC Brivo Mobile C-Arm X-Ray Products are designed to provide digital spot film imaging and fluoroscopic image guidance for all adult and pediatric populations for orientations between patient anatomy and surgical instruments. The product is used for general surgical applications and musculoskeletal procedures to visualize, for example, implant localizations or needle positions for aspirations,injections or biopsy.
- Status
- Terminated
- Initiated
- 2013-09-18
- Posted
- 2013-10-24
- Terminated
- 2014-01-27
- Reason
- GE is recalling certain lots of GE Brivo OEC715/785/865 C-Arm due to the potential for the wires in the high voltage cable bundle to fail, resulting in loss of system functionality.
- Root cause
- Device Design
- Action
- GE Healthcare sent an Urgent Medical Device Correction letter dated September 18,2013, to all affected customers to inform them of the issue and provide instructions for immediate mitigation of the issue. Customers were instructed to ensure that all potential users in their facility were made aware of this safety notification and the recommended actions. Customers with questions were instructed to call 1-800-874-7378 option 8. For questions regarding this recall call 801-295-4952.
- Quantity
- 29
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to China, India, Japan, Malaysia, Australia, Ireland, France, Germany, Spain, Hungary, Belgium, and Switzerland.
- Lot, serial or model codes
- GE Brivo OEC715: Serial Numbers B2S13001 ~ B2S13010. GE Brivo OEC785: Serial Numbers B3S13005 ~ B3S13008. GE Brivo OEC865: Serial Numbers B4S13001 ~ B4S13015
- 510(k) numbers
- ["K123603"]
- PMA numbers
- not on file
- Product code
- OXO
- Firm FEI
- 3002175190
- Firm city
- Beijing
- Firm state
- not on file
- Firm country
- China
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28