FDA Medical Device Recalls (openFDA)
Smiths Medical MD, Inc.: Locator-Wand Cover that is sold in a shelf carton (unit packaging) separate to the sterile tray containing and identified as PORT-A-CATH II Fluoro-Free, Low Profile Venous Access System, PolyFlow Polyurethane Catheter, 1.9mm O.D. tray. REF 21-4683-24. Contents: Portal with ULTRA-LOCK Connector, Radiopaque PolyFlow polyurethane Catheter with CATH-FINDER Sensor Wire, Locator-Wand cover, Nonvented Luer Cap, 20G PORT-A-CATH needle, 18G Blunt needle, 20G GRIPPER PLUS non-Coring Safety needle, 18G Extra Thin Wall Introducer Needle, 6 Fr introducer, 2 ''J'' Guidewires, Syringes, Vein Pick, Tunneling Tool, 20G 90-degree PORT-A-CATH needle, 3 Point-Lok Sharps Safety Devices, Product Information. EtO Sterilized. Single Use. Deltec, Inc., St. Paul, MN 55112 U.S.A.
- Recall number
- Z-0066-2007
- Recalling firm
- Smiths Medical MD, Inc.
- Product
- Locator-Wand Cover that is sold in a shelf carton (unit packaging) separate to the sterile tray containing and identified as PORT-A-CATH II Fluoro-Free, Low Profile Venous Access System, PolyFlow Polyurethane Catheter, 1.9mm O.D. tray. REF 21-4683-24. Contents: Portal with ULTRA-LOCK Connector, Radiopaque PolyFlow polyurethane Catheter with CATH-FINDER Sensor Wire, Locator-Wand cover, Nonvented Luer Cap, 20G PORT-A-CATH needle, 18G Blunt needle, 20G GRIPPER PLUS non-Coring Safety needle, 18G Extra Thin Wall Introducer Needle, 6 Fr introducer, 2 ''J'' Guidewires, Syringes, Vein Pick, Tunneling Tool, 20G 90-degree PORT-A-CATH needle, 3 Point-Lok Sharps Safety Devices, Product Information. EtO Sterilized. Single Use. Deltec, Inc., St. Paul, MN 55112 U.S.A.
- Status
- Terminated
- Initiated
- 2006-09-27
- Posted
- 2006-10-24
- Terminated
- 2006-12-30
- Reason
- Smiths Medical became aware that product returned to them under recall # Z-0800-06/ Z-0807-06 was inadvertently distributed to consignees after the recall was initiated 03/29/2006. The product was originally recalled because Locator Wand covers, which are supplied on the outside of PORT-A-CATH and P.A.S. PORT sterile trays, may have an insufficient seal and product sterility could be compromised.
- Root cause
- Other
- Action
- An Urgent Product Safety & Recall Notification, dated 09/27/06, was sent to customers via facsimile and phone calls were made. The letter describes the issue and customers are given 3 options to include: use only the sterile tray and discard the wand cover, remove and discard Locator-Wand Cover pouches and request replacement Locator-Wand Covers, or return the entire device for a credit or replacement.
- Quantity
- 1 tray
- Distribution
- CA, FL, MA, MD, OH, OK, TX
- Lot, serial or model codes
- Lot number M42880
- 510(k) numbers
- ["K871132","K931073","K942024"]
- PMA numbers
- not on file
- Product code
- MMP
- Firm FEI
- 2183502
- Firm city
- Saint Paul
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28