Skip to the content

FDA Medical Device Recalls (openFDA)

Smiths Medical MD, Inc.: Locator-Wand Cover that is sold in a shelf carton (unit packaging) separate to the sterile tray containing and identified as PORT-A-CATH II Fluoro-Free, Low Profile Venous Access System, PolyFlow Polyurethane Catheter, 1.9mm O.D. tray. REF 21-4683-24. Contents: Portal with ULTRA-LOCK Connector, Radiopaque PolyFlow polyurethane Catheter with CATH-FINDER Sensor Wire, Locator-Wand cover, Nonvented Luer Cap, 20G PORT-A-CATH needle, 18G Blunt needle, 20G GRIPPER PLUS non-Coring Safety needle, 18G Extra Thin Wall Introducer Needle, 6 Fr introducer, 2 ''J'' Guidewires, Syringes, Vein Pick, Tunneling Tool, 20G 90-degree PORT-A-CATH needle, 3 Point-Lok Sharps Safety Devices, Product Information. EtO Sterilized. Single Use. Deltec, Inc., St. Paul, MN 55112 U.S.A.

Recall number
Z-0066-2007
Recalling firm
Smiths Medical MD, Inc.
Product
Locator-Wand Cover that is sold in a shelf carton (unit packaging) separate to the sterile tray containing and identified as PORT-A-CATH II Fluoro-Free, Low Profile Venous Access System, PolyFlow Polyurethane Catheter, 1.9mm O.D. tray. REF 21-4683-24. Contents: Portal with ULTRA-LOCK Connector, Radiopaque PolyFlow polyurethane Catheter with CATH-FINDER Sensor Wire, Locator-Wand cover, Nonvented Luer Cap, 20G PORT-A-CATH needle, 18G Blunt needle, 20G GRIPPER PLUS non-Coring Safety needle, 18G Extra Thin Wall Introducer Needle, 6 Fr introducer, 2 ''J'' Guidewires, Syringes, Vein Pick, Tunneling Tool, 20G 90-degree PORT-A-CATH needle, 3 Point-Lok Sharps Safety Devices, Product Information. EtO Sterilized. Single Use. Deltec, Inc., St. Paul, MN 55112 U.S.A.
Status
Terminated
Initiated
2006-09-27
Posted
2006-10-24
Terminated
2006-12-30
Reason
Smiths Medical became aware that product returned to them under recall # Z-0800-06/ Z-0807-06 was inadvertently distributed to consignees after the recall was initiated 03/29/2006. The product was originally recalled because Locator Wand covers, which are supplied on the outside of PORT-A-CATH and P.A.S. PORT sterile trays, may have an insufficient seal and product sterility could be compromised.
Root cause
Other
Action
An Urgent Product Safety & Recall Notification, dated 09/27/06, was sent to customers via facsimile and phone calls were made. The letter describes the issue and customers are given 3 options to include: use only the sterile tray and discard the wand cover, remove and discard Locator-Wand Cover pouches and request replacement Locator-Wand Covers, or return the entire device for a credit or replacement.
Quantity
1 tray
Distribution
CA, FL, MA, MD, OH, OK, TX
Lot, serial or model codes
Lot number M42880
510(k) numbers
["K871132","K931073","K942024"]
PMA numbers
not on file
Product code
MMP
Firm FEI
2183502
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register