Skip to the content

FDA Medical Device Recalls (openFDA)

Encore Medical, Lp: Cobalt MV Bone Cement 40GM, REF 402438, QTY 1, STERILE, Rx Only, Mfgr: BIOMET ORTHOPEDICS Cobalt MV and HV Bone Cement is indicated for use as bone cement in arthroplastic procedures of the hip, knee and other joints to fix plastic and metal prosthetic parts to living bone when reconstructions is necessary because of osteoarthritis, rheumatoid arthritis, traumatic arthritis, avascular necrosis, nonunion of fractures of the neck of the femur, sickle cell anemia osteoporosis, secondary severe joint destruction following trauma or other conditions (also far fixation of unstable fractures in metastatic malignancies), and revision of previous arthroplasty procedures

Recall number
Z-0065-2018
Recalling firm
Encore Medical, Lp
Product
Cobalt MV Bone Cement 40GM, REF 402438, QTY 1, STERILE, Rx Only, Mfgr: BIOMET ORTHOPEDICS Cobalt MV and HV Bone Cement is indicated for use as bone cement in arthroplastic procedures of the hip, knee and other joints to fix plastic and metal prosthetic parts to living bone when reconstructions is necessary because of osteoarthritis, rheumatoid arthritis, traumatic arthritis, avascular necrosis, nonunion of fractures of the neck of the femur, sickle cell anemia osteoporosis, secondary severe joint destruction following trauma or other conditions (also far fixation of unstable fractures in metastatic malignancies), and revision of previous arthroplasty procedures
Status
Open, Classified
Initiated
2017-06-26
Posted
2017-11-06
Terminated
not on file
Reason
Loss of the seal on the sterile Tyvek packaging used with this Cobalt Bone Cement.
Root cause
Vendor change control
Action
Zimmer Biomet sent an Urgent Medical Device Recall Notice dated April 2017 to all affected customers. The firm initiated their recall to their distributors on 06/26/2017 requesting that they destroy any product on hand. The firm then expanded their recall on 09/14/2017, beginning their notification by email and following with letters to hospitals delivered on 09/21 and 09/22/2017. The firm requested the following actions: "1. Complete the Acknowledgement/Questionnaire provided with this lot specific field safety notice and return to Stericycle by email to DJO5504@stericycle.com or by fax to 866-608-3939. If you have any questions or issues related to sending this information, please call 877-551-7153. 2. Quarantine this affected product. Arrangements will be made to finalize the handling of the product as well as coordinate the replacement product where applicable. 3. If packaging issues are found, please contact DJO Surgical Customer Service to initiate a product complaint." Customers with questions were instructed to call 512-834-6255.
Quantity
840 units
Distribution
Nationwide Distribution
Lot, serial or model codes
Lot Numbers: 507830, 871270, 959670
510(k) numbers
["K091608"]
PMA numbers
not on file
Product code
LOD
Firm FEI
1000116912
Firm city
Austin
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register