Skip to the content

FDA Medical Device Recalls (openFDA)

Winco Mfg., LLC: TransMotion Medical, TMM3 Series Video-Fluoroscopy Swallow Study Stretcher-Chair. Intended use is to provide the caregiver a single platform for receiving, transporting within a facility, performing procedures, recovery and discharge of patients in a monitored outpatient environment.

Recall number
Z-0064-2017
Recalling firm
Winco Mfg., LLC
Product
TransMotion Medical, TMM3 Series Video-Fluoroscopy Swallow Study Stretcher-Chair. Intended use is to provide the caregiver a single platform for receiving, transporting within a facility, performing procedures, recovery and discharge of patients in a monitored outpatient environment.
Status
Terminated
Initiated
2016-06-29
Posted
not on file
Terminated
2017-09-27
Reason
Non-conforming component caused arms of TMM4 Stretcher chair to no longer meet design specifications.
Root cause
Nonconforming Material/Component
Action
Customers were sent an Urgent: Medical Device Recall notification on 7/12/2016 for the specific models of the TransMotion (TMM) Stretcher chairs. The letter explained the reason for the recall and asked customers to take action. Customers are asked to follow the instruction in the Operator Manual for proper egress. Patients are not allowed to egress from the chair without assistance nor should they be left unattended. The chairs will be repaired on site; and, removal of the chair from service is not required with proper supervision and use. Customers are asked to complete the instructions for the acknowledgement form and return. The firm will contact the customers to arrange for repair. Questions regarding the field correction should be directed to the Customer Care department at 800-237-3377 or 352-854-2929 or e-mail QA-RA@wincomfg.com
Quantity
264 Stretcher Chairs
Distribution
Worldwide Distribution -- USA, including AR, CA, CO, CT, FL, GA, IL, KS, LA, MD, MI, MN, MO, MS, NC, NJ, NM, NV, NY, OH, OK, PA, RI, TN, TX, VA, WA, WI, and Puerto Rico; and the countries of Canada, Australia, Italy, Panama, Switzerland and Taiwan.
Lot, serial or model codes
All identified serial numbers.
510(k) numbers
not on file
PMA numbers
not on file
Product code
FPP
Firm FEI
1027229
Firm city
Ocala
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register