Ortho Clinical Diagnostics Inc: VITROS Immunodiagnostic Products - TSH Reagent Pack (10758750000227) 1912997

Recall number
Z-0061-2019
Recalling firm
Ortho Clinical Diagnostics Inc
Product
VITROS Immunodiagnostic Products - TSH Reagent Pack (10758750000227) 1912997
Status
Open, Classified
Initiated
2018-07-12
Posted
not on file
Terminated
not on file
Reason
The recalling firm determined that biased results may occur for specific VITROS Immunodiagnostic Products (MicroWell Assays) at biotin concentrations which are lower than indicated in the current Instructions For Use (IFU). Patients who are taking biotin supplements could potentially have biased sample results for affected assays at biotin concentrations lower than indicated in the current IFU.
Root cause
Device Design
Action
On 12 July 2018, a customer letter was sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail to all customers who could process the affected products (i.e., VITROS 3600, 5600 and/or ECi/ECiQ Systems) to inform them of the potential interference of biotin at levels lower than stated in the IFUs and that Ortho is in the process of revising the IFUs for the affected products. Ortho recommended customers follow normal laboratory procedures for troubleshooting of samples containing other assay interferences and it was acceptable to continue using the affected products. If you have questions, please contact Ortho Care Technical Solutions Center at 1-800-421-3311.
Quantity
68230
Distribution
Worldwide Distribution - US Nationwide distribution, and countries of Australia, Bermuda, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Singapore, United Kingdom, France, Germany, Italy, Spain, Portugal, Poland, Russia, Denmark, Norway, Sweden, Belgium, and Netherlands.
Lot, serial or model codes
5535 5540 5550 5560 5570 5580 5590 5600 5610 5615 5620 5630 5640 5650 5655 5660 5665 5670 5675 5678 5680 5685 5690
510(k) numbers
["K081543"]
PMA numbers
not on file
Product code
JLW
Firm FEI
2250051
Firm city
Raritan
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register