FDA Medical Device Recalls (openFDA)
DePuy Orthopaedics, Inc.: Charnley PN Retract & Handle, REF 962004000, Non-Sterile-DuPuy International Ltd. The Charnley Pin Retractor and Handle Set is used to facilitate the insertion and extraction of two pins into and out of bone in the incision area, thereby holding the incision soft tissue open during the surgical procedure.
- Recall number
- Z-0061-2015
- Recalling firm
- DePuy Orthopaedics, Inc.
- Product
- Charnley PN Retract & Handle, REF 962004000, Non-Sterile-DuPuy International Ltd. The Charnley Pin Retractor and Handle Set is used to facilitate the insertion and extraction of two pins into and out of bone in the incision area, thereby holding the incision soft tissue open during the surgical procedure.
- Status
- Terminated
- Initiated
- 2014-08-27
- Posted
- 2014-10-09
- Terminated
- 2015-11-17
- Reason
- The small extraction peg of the Charnley Pin Retractor and Handle set is breaking from the pin.
- Root cause
- Mixed-up of materials/components
- Action
- The firm, Depuy Synthes, sent an "URGENT INFORMATION - RECALL NOTICE" dated August 27, 2014 to their consignees. The recall notification included a description of the reason for recall, affected product, clinical implications, and instructions for responding to the formal recall notification. The customers were instructed to cease using the affected components immediately, quarantine all affected components in a manner that ensures the affected components are not used; return affected devices immediately to your DePuy; if any affected product has been forwarded to another facility, contact that facility immediately to communicate recall and arrange return; maintain awareness and copy of this notice, and complete and return the Reconciliation forms via fax to 574-371-4939 or emailed to kearle@its.jnj.com. For questions about the device recall information provided, please contact Recall Coordinator, at 574-371-4917 (M-F; 8 a.m. - 5 p.m. EDT) or email Kearle@its.jnj.com. For product-related questions, please contact your local DePuy Orthopaedics, Inc. Sales Consultant. For clinical questions from hospitals/user facilities, please contact DePuy Orthopaedics, Inc. Scientific Information Office at 1-888-554-2482 (M-F; 8 a.m. - 5 p.m. EDT).
- Quantity
- 65 units
- Distribution
- Worldwide Distribution: US (Nationwide) including states of: AR, CA, CO, CT, IL, MA, NC, OH, PA, TN, TX, UT, VA, WA, WI; and Internationally to: AUSTRALIA, CHILE, UK, INDIA, IRELAND, JAPAN, NEW ZEALAND, RUSSIA, SINGAPORE, and SAUDI ARABIA.
- Lot, serial or model codes
- Barcode: 10603295240532 CATALOG NO. : 962004000 Lot # OSA-149328 Lot # OSA-181610 Lot # OSA-161157
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- MDM
- Firm FEI
- 1818910
- Firm city
- Warsaw
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28