Skip to the content

FDA Medical Device Recalls (openFDA)

DePuy Orthopaedics, Inc.: Charnley PN Retract & Handle, REF 962004000, Non-Sterile-DuPuy International Ltd. The Charnley Pin Retractor and Handle Set is used to facilitate the insertion and extraction of two pins into and out of bone in the incision area, thereby holding the incision soft tissue open during the surgical procedure.

Recall number
Z-0061-2015
Recalling firm
DePuy Orthopaedics, Inc.
Product
Charnley PN Retract & Handle, REF 962004000, Non-Sterile-DuPuy International Ltd. The Charnley Pin Retractor and Handle Set is used to facilitate the insertion and extraction of two pins into and out of bone in the incision area, thereby holding the incision soft tissue open during the surgical procedure.
Status
Terminated
Initiated
2014-08-27
Posted
2014-10-09
Terminated
2015-11-17
Reason
The small extraction peg of the Charnley Pin Retractor and Handle set is breaking from the pin.
Root cause
Mixed-up of materials/components
Action
The firm, Depuy Synthes, sent an "URGENT INFORMATION - RECALL NOTICE" dated August 27, 2014 to their consignees. The recall notification included a description of the reason for recall, affected product, clinical implications, and instructions for responding to the formal recall notification. The customers were instructed to cease using the affected components immediately, quarantine all affected components in a manner that ensures the affected components are not used; return affected devices immediately to your DePuy; if any affected product has been forwarded to another facility, contact that facility immediately to communicate recall and arrange return; maintain awareness and copy of this notice, and complete and return the Reconciliation forms via fax to 574-371-4939 or emailed to kearle@its.jnj.com. For questions about the device recall information provided, please contact Recall Coordinator, at 574-371-4917 (M-F; 8 a.m. - 5 p.m. EDT) or email Kearle@its.jnj.com. For product-related questions, please contact your local DePuy Orthopaedics, Inc. Sales Consultant. For clinical questions from hospitals/user facilities, please contact DePuy Orthopaedics, Inc. Scientific Information Office at 1-888-554-2482 (M-F; 8 a.m. - 5 p.m. EDT).
Quantity
65 units
Distribution
Worldwide Distribution: US (Nationwide) including states of: AR, CA, CO, CT, IL, MA, NC, OH, PA, TN, TX, UT, VA, WA, WI; and Internationally to: AUSTRALIA, CHILE, UK, INDIA, IRELAND, JAPAN, NEW ZEALAND, RUSSIA, SINGAPORE, and SAUDI ARABIA.
Lot, serial or model codes
Barcode: 10603295240532 CATALOG NO. : 962004000  Lot # OSA-149328 Lot # OSA-181610  Lot # OSA-161157
510(k) numbers
not on file
PMA numbers
not on file
Product code
MDM
Firm FEI
1818910
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register