Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer, Inc.: CPT Modular Stem, Item No. 00-8011-001-00; Lot 61947708. CPT stems are primary cemented femoral hip stems used by surgeons in total hip arthroplasty.

Recall number
Z-0060-2013
Recalling firm
Zimmer, Inc.
Product
CPT Modular Stem, Item No. 00-8011-001-00; Lot 61947708. CPT stems are primary cemented femoral hip stems used by surgeons in total hip arthroplasty.
Status
Terminated
Initiated
2012-09-13
Posted
2012-10-15
Terminated
2014-01-15
Reason
Zimmer Inc. is initiating a removal of the CPT Modular Stem (item 00-8011-001-00), CPT Stem Petite (item 00-8011-000-05; ) and CPT Femoral Stem 12/14 Neck Taper Extended Offset,(item 00-8114-001-10 ) because devices were packaged in incorrect cartons and distributed internationally and domestically. Zimmer received five complaints of this packaging issue.
Root cause
Packaging change control
Action
Zimmer Inc. sent an "URGENT MEDICAL DEVICE RECALL-LOT SPECIFIC" letter dated September 13, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Consignees were asked to return all recalled product with a completed Inventory Return Certification Form. Contact the firm at 1-800-348-2759 for questions regarding this notice.
Quantity
59 both units
Distribution
Worldwide Distribution-USA (nationwide) including the states of Indiana and Oklahoma and the countries of Australia, Thailand, Malaysia, Germany, India, Bulgaria, Egypt, UK, Russian FED, Slovak, and Brazil.
Lot, serial or model codes
Lot 61947708
510(k) numbers
not on file
PMA numbers
not on file
Product code
not on file
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register