Cardinal Health: Presource Custom Sterile Packs containing the Cardiovascular Split Drape II; the custom sterile packs were packaged with the bulk, non-sterile drapes and then the complete custom pack was sterilized; packaged by Cardinal Health Medical Products and Services, McGaw Park, IL 60085; identified by the following catalog numbers: a) SVC13AMEMK - Sterile Open Heart Adult Pk; b) SCV31OHMPJ - Sterile Open Heart w Drapes Pk; c) SVC42OPVCG - Sterile Open Heart Adult Pack; d) SCV32CPMUO - Sterile CAB 146060 Pack; e) SCV34OHTMJ - Sterile Basin Pack; f) SCV31LVLUM - Sterile Liver Transplant Pack; g) SCV19CVHHF - Sterile Cardiovascular Pack; h) SCV11ADLLD - Sterile Open Heart/Adult; i) SCV92CBHRB - Sterile CABG Pack; j) SCV92CASIA - Sterile Cardiovascular Pack; k) SBA32GSMUL - Sterile General Surgery Pack; l) SCV30OHKWA - Strl **** Open Heart Pack; m) SBA31LVLUM - Sterile Liver Transplant Pack; n) SBA35LTBLQ - Sterile Lung Transplant Pack; o) SCV44CVDEB - Sterile CV Pack; p) SCV22OHMES - Open Heart Pack; q) SCV23AOSFH - Open Heart Part 3; r) SCV43OHHRF - Open Heart; s) SCV24CAGME - CABG Drape Pack; t) SAN12CCKGB - Cardiac Pack; u) SBA13BRBSE - Breast Reduction; v) SCV11CBNYP - CABG G 3; w) SCV11CESLB - C.V. Drape Pack; x) SCV11CGCPR - CABG; y) SCV11CGXXB - CABG Pack; z) SCV11CJICL - Cardiac Major; aa) SCV11CTMUK - C T Set Up; bb) SCV11DPMYG - **** S Drape Pack; cc) SCV11HEMMI - Open Heart A B; dd) SCV11MRCPE - Major Cardiac; ee) SCV11OCNYN - OH CABG; ff) SCV11OHLD2 - Open Heart A&B Pack; gg) SCV11OHNMC - Open Heart; hh) SCV11OPLKC - O P CABG; ii) SCV11VASLB - C V Valve Drape; jj) SCV11VLNYM - G 3 Valve; kk) SCV12OHYHY - Graft and Linen; ll) SCV13AABSV - Aortic Aneurysm; mm) SCV13AMEMJ - Open Heart Pack B; nn) SCV13BBEMC - Cardio-Thoracic Aux; oo) SCV13CB767 - CABG; pp) SCV13OHCAI - Open Heart Pk; qq) SCV13OOBIG - Open Heart Drape; rr) SCV14O1BGO - Open Heart Pack A; ss) SCV22OHPGA - Open Heart Pack; tt) SCV22OOMSE - Open Heart Preferred; uu) SCV24OH20A - Open Heart Drape Pack; vv) SCV43DREJI - Open Heart Drape Pack; ww) SCV69OH17E - Open Heart; xx) SCV59HVSGJ - Heart Valve Pack; yy) SCV59OHNVM - Open Heart Pack; zz) SCV53HTDMG - ***** Open Heart; aaa) SCV56CH11A - Open Heart Pack; bbb) SCV59MJTCW - Open Heart Pack; ccc) SCV52OH97F - Open Heart Pack; ddd) SCV41OHHIE - Open Heart Pack; eee) SCV52HTSJE - Open Heart Pack; fff) SCV59OHMHV - Basic Heart Pack; ggg) SCV11MJTJI - Open Heart

Recall number
Z-0060-2007
Recalling firm
Cardinal Health
Product
Presource Custom Sterile Packs containing the Cardiovascular Split Drape II; the custom sterile packs were packaged with the bulk, non-sterile drapes and then the complete custom pack was sterilized; packaged by Cardinal Health Medical Products and Services, McGaw Park, IL 60085; identified by the following catalog numbers: a) SVC13AMEMK - Sterile Open Heart Adult Pk; b) SCV31OHMPJ - Sterile Open Heart w Drapes Pk; c) SVC42OPVCG - Sterile Open Heart Adult Pack; d) SCV32CPMUO - Sterile CAB 146060 Pack; e) SCV34OHTMJ - Sterile Basin Pack; f) SCV31LVLUM - Sterile Liver Transplant Pack; g) SCV19CVHHF - Sterile Cardiovascular Pack; h) SCV11ADLLD - Sterile Open Heart/Adult; i) SCV92CBHRB - Sterile CABG Pack; j) SCV92CASIA - Sterile Cardiovascular Pack; k) SBA32GSMUL - Sterile General Surgery Pack; l) SCV30OHKWA - Strl **** Open Heart Pack; m) SBA31LVLUM - Sterile Liver Transplant Pack; n) SBA35LTBLQ - Sterile Lung Transplant Pack; o) SCV44CVDEB - Sterile CV Pack; p) SCV22OHMES - Open Heart Pack; q) SCV23AOSFH - Open Heart Part 3; r) SCV43OHHRF - Open Heart; s) SCV24CAGME - CABG Drape Pack; t) SAN12CCKGB - Cardiac Pack; u) SBA13BRBSE - Breast Reduction; v) SCV11CBNYP - CABG G 3; w) SCV11CESLB - C.V. Drape Pack; x) SCV11CGCPR - CABG; y) SCV11CGXXB - CABG Pack; z) SCV11CJICL - Cardiac Major; aa) SCV11CTMUK - C T Set Up; bb) SCV11DPMYG - **** S Drape Pack; cc) SCV11HEMMI - Open Heart A B; dd) SCV11MRCPE - Major Cardiac; ee) SCV11OCNYN - OH CABG; ff) SCV11OHLD2 - Open Heart A&B Pack; gg) SCV11OHNMC - Open Heart; hh) SCV11OPLKC - O P CABG; ii) SCV11VASLB - C V Valve Drape; jj) SCV11VLNYM - G 3 Valve; kk) SCV12OHYHY - Graft and Linen; ll) SCV13AABSV - Aortic Aneurysm; mm) SCV13AMEMJ - Open Heart Pack B; nn) SCV13BBEMC - Cardio-Thoracic Aux; oo) SCV13CB767 - CABG; pp) SCV13OHCAI - Open Heart Pk; qq) SCV13OOBIG - Open Heart Drape; rr) SCV14O1BGO - Open Heart Pack A; ss) SCV22OHPGA - Open Heart Pack; tt) SCV22OOMSE - Open Heart Preferred; uu) SCV24OH20A - Open Heart Drape Pack; vv) SCV43DREJI - Open Heart Drape Pack; ww) SCV69OH17E - Open Heart; xx) SCV59HVSGJ - Heart Valve Pack; yy) SCV59OHNVM - Open Heart Pack; zz) SCV53HTDMG - ***** Open Heart; aaa) SCV56CH11A - Open Heart Pack; bbb) SCV59MJTCW - Open Heart Pack; ccc) SCV52OH97F - Open Heart Pack; ddd) SCV41OHHIE - Open Heart Pack; eee) SCV52HTSJE - Open Heart Pack; fff) SCV59OHMHV - Basic Heart Pack; ggg) SCV11MJTJI - Open Heart
Status
Terminated
Initiated
2006-09-27
Posted
2006-10-21
Terminated
2008-05-09
Reason
The cardiovascular drape may tear and/or fray at the reinforced fenestrated trough area during use.
Root cause
Other
Action
Cardinal Health sent separate recall letters dated 9/27/06 via UPS next day delivery to those accounts who ordered the drapes alone and those accounts who ordered custom sterile packs and PBDS modules, informing them that the cardiovascular split drapes may tear and/or fray at the reinforced fenestrated trough area during use. The letters listed the catalog numbers and affected lot numbers of the sterile drapes, and the letters sent to the pack and module customers were accompanied by a listing of catalog numbers and work order numbers for the affected packs ordered by each account. The accounts were requested to inspect their inventory to determine if they had any of the affected product on hand; segregate the affected inventories; complete and return via fax at 847-689-9101, the enclosed acknowledgment form, indicating the number of drapes in inventory being destroyed; and contact the El Paso Customer Service Center at 915-779-3681 to order replacement product. Distributors were requested to notify their customers of the recall. An additional letter dated 10/25/06 was sent via UPS overnight delivery to 49 of the accounts, expanding the recall to include lot 06HRD110 of catalog #9158, Convertors Tiburon Cardiovascular Split Drape II, Sterile. The instructions remained the same as those in the 9/27/06 letter.
Quantity
3,533 packs
Distribution
Worldwide Distribution- USA and countries of Kuwait and Australia
Lot, serial or model codes
cat #SVC13AMEMK, order #777594;  cat #SCV31OHMPJ, order #773933;  cat #SVC42OPVCG, order #774231;  cat #SCV32CPMUO, order #774845;  cat #SCV34OHTMJ, order #775103;  cat #SCV31LVLUM, order #775646;  cat #SCV19CVHHF, order #776102;  cat #SCV11ADLLD, order #776475;  cat #SCV92CBHRB, order #776785;  cat #SCV92CASIA, order #776792;  cat #SBA32GSMUL, order #777047;  cat #SCV30OHKWA, order #777171;  cat #SBA31LVLUM, order #777288;  cat #SBA35LTBLQ, order #777599;  cat #SCV44CVDEB, order #780376;  cat #SCV22OHMES, order #731809;  cat #SCV23AOSFH, order #737659;  cat #SCV43OHHRF, order #739410;  cat #SCV24CAGME, order #740765;  cat #SAN12CCKGB, order #732285;  cat #SBA13BRBSE, order #731714;  cat #SCV11CBNYP, order #732629;  cat #SCV11CESLB, order #734779;  cat #SCV11CGCPR, order #732923;  cat #SCV11CGXXB, order #735832;  cat #SCV11CJICL, order #735189;  cat #SCV11CTMUK, order #734463;  cat #SCV11DPMYG, order #732152;  cat #SCV11HEMMI, order #733639;  cat #SCV11MRCPE, order #733157;  cat #SCV11OCNYN, order #732625;  cat #SCV11OHLD2, order #732191;  cat #SCV11OHNMC, order #732446;  cat #SCV11OPLKC, order #735134;  cat #SCV11VASLB, order #737823;  cat #SCV11VLNYM, order #732656;  cat #SCV12OHYHY, order #735934, 734739, 732575;  cat #SCV13AABSV, order #732913;  cat #SCV13AMEMJ, order #732970;  cat #SCV13BBEMC, order #732070;  cat #SCV13CB767, order #732869;  cat #SCV13OHCAI, order #733777, 731664;  cat #SCV13OOBIG, order #734407;  cat #SCV14O1BGO, order #738338;  cat #SCV22OHPGA, order #735643;  cat #SCV22OOMSE, order #738400;  cat #SCV24OH20A, order #735637;  cat #SCV43DREJI, order #741763;  cat #SCV69OH17E, order #975704;  cat #SCV59HVSGJ, order #975779;  cat #SCV59OHNVM, order #976192;  cat #SCV53HTDMG, order #976853;  cat #SCV56CH11A, order #977429;  cat #SCV59MJTCW, order #977839, 978730;  cat #SCV52OH97F, order #978285;  cat #SCV41OHHIE, order #978409;  cat #SCV52HTSJE, order #978410;  cat #SCV59OHMHV, order #978523; cat #SCV11MJTJI, order 737814
510(k) numbers
not on file
PMA numbers
not on file
Product code
LRO
Firm FEI
3001236905
Firm city
Park
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register