FDA Medical Device Recalls (openFDA)
Biosense Webster, Inc.: THERMOCOOL SF NAV Catheters Model Numbers - BNI35FJCT, BNI35DFCT, D-131503-S, D-131504-S
- Recall number
- Z-0058-2018
- Recalling firm
- Biosense Webster, Inc.
- Product
- THERMOCOOL SF NAV Catheters Model Numbers - BNI35FJCT, BNI35DFCT, D-131503-S, D-131504-S
- Status
- Terminated
- Initiated
- 2017-09-06
- Posted
- not on file
- Terminated
- 2019-11-19
- Reason
- Biosense Webster, Inc. has recently received an increased number of complaints related to the display of Alert 402 on the CARTO 3 System for certain lots of THERMOCOOL brand catheters. Alert 402 implies a "Map: magnetic distortion" when connected to CARTO 3 System. This issue may subsequently lead the physician to ablate in an unintended area when delivering RF energy.
- Root cause
- Under Investigation by firm
- Action
- All US and OUS customers, who have purchased the lots identified as part of this recall will be notified via customer letters in order to return the product to BWI facility where it will be segregated for further disposition. This product recall applies to hospitals (or direct customers) and distributors who have purchased and/or received the THERMOCOOL Catheters identified as part of this recall . One notification package which includes the Customer Notification Letter, Business Reply Form (B RF) and pre-paid return label will be distributed to the customers subject to this recall via a 3rd party vendor (Stericycle) for USA. OUS customer communication will occur in accordance with local regulations.
- Quantity
- 329 units
- Distribution
- US, Japan, Canada, Australia, New Zealand, Brazil, Croatia, Korea, Austria, Colombia, Italy, Portugal, Argentina, South Africa, Belgium, France, Germany, Ireland, Poland, Netherlands, Sweden, Denmark, Finland, Switzerland, Great Britain, Hungary, Czech Republic, China
- Lot, serial or model codes
- 17637495L 17638389L 17639706L 17639707L 17639708L 17640039L 17640494L 17640952L 17640040L 17640742L 17640953L 17640041L 17640743L 17640954L 17640490L 17640744L 17643130L 17638118L 17638387L 17638388L
- 510(k) numbers
- not on file
- PMA numbers
- ["P990071","P030031S025","P030031S034"]
- Product code
- LPB
- Firm FEI
- 3003113629
- Firm city
- Irwindale
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28