Skip to the content

FDA Medical Device Recalls (openFDA)

Thoratec Corp: HeartMate II System Controller, a component of the HeartMate II Left Ventricular Device System (LVAS), distributed both as stand alone devices and as components of the LVAS Implant kit. Manufactured/Distributed by Thoratec Corporation, Pleasanton, CA The HeartMate II Left Ventricular Assist System (LVAS) consists of an implantable blood pump connected to an eternal system controller by a percutaneous lead. The external controller is powered bye either batteries or a power supply that connects to AC Main power. The device is intended for use as a bridge to transplantation in cardiac transplant candidates at risk of imminent death from non-reversible left ventricular failure. It is also for use in patients with New York Heart Association class IIIB or IV end-stage left ventricular

Recall number
Z-0058-2011
Recalling firm
Thoratec Corp
Product
HeartMate II System Controller, a component of the HeartMate II Left Ventricular Device System (LVAS), distributed both as stand alone devices and as components of the LVAS Implant kit. Manufactured/Distributed by Thoratec Corporation, Pleasanton, CA The HeartMate II Left Ventricular Assist System (LVAS) consists of an implantable blood pump connected to an eternal system controller by a percutaneous lead. The external controller is powered bye either batteries or a power supply that connects to AC Main power. The device is intended for use as a bridge to transplantation in cardiac transplant candidates at risk of imminent death from non-reversible left ventricular failure. It is also for use in patients with New York Heart Association class IIIB or IV end-stage left ventricular
Status
Terminated
Initiated
2010-09-01
Posted
2010-10-13
Terminated
2011-04-07
Reason
Connection failure-- product may malfunction because of bent pins or broken wires in power cord connection
Root cause
Device Design
Action
Thoratec Corporation sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated August 30, 2010, to all customers. The letter described the affected product, description and symptoms of problem, and immediate action to be taken by the customers. The customers were instructed to: 1) contact all ongoing patients using the affected Systems Controllers (serial numbers starting with "EPC") to determine if they are experiencing any unexplained intermittent alarms or have any bent pins in the connectors between the System Controller and the power sources, if so, have them return to the hospital so that their power leads can be examined. If further assistance is needed with the evaluation, please contact Thoratec Technical Service (800-456-1477). 2) review the attached addendum to the HeartMate II labeling with all VAD clinicians and ongoing HeartMate II LVAS patients, and 3) complete and sign the attached Acknowledgment Form via fax to Thoratec Regulatory Affairs at (952) 847-8571. If you have any questions, please contact 925-847-8600.
Quantity
8965 units
Distribution
Worldwide distribution: USA including states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX , UT, VA, WA, and WI; and countries including: Austria, Belgium, Canada, Czech Republic, Denmark, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Lithuania, Netherlands, Singapore, Slovenia, Sweden, Switzerland, and the UK.
Lot, serial or model codes
All serial numbers starting with EPC prefix
510(k) numbers
not on file
PMA numbers
["P060040"]
Product code
DSQ
Firm FEI
2916596
Firm city
Pleasanton
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register