Skip to the content

FDA Medical Device Recalls (openFDA)

Unomedical: Hospitak Infant Manual Pulmonary Resuscitator with reservoir tube, face mask and Pop-Off 40 cm H2O Elbow. Foreign distribution only: Product codes: 734-E, 734MM.

Recall number
Z-0058-2010
Recalling firm
Unomedical
Product
Hospitak Infant Manual Pulmonary Resuscitator with reservoir tube, face mask and Pop-Off 40 cm H2O Elbow. Foreign distribution only: Product codes: 734-E, 734MM.
Status
Terminated
Initiated
2009-10-06
Posted
2009-11-04
Terminated
2010-08-10
Reason
In some Manual Pulmonary Resuscitator units the duckbill valve and retention ring inside the patient valve housing may become dislodged and fall into the MPR bag/body. If this occurs, the MPR may lose its ability to generate the positive pressure necessary necessary to ventilate the patient.
Root cause
Device Design
Action
Unomedical issued Press on October 6, 2009. Letters were sent to customers on October 7, 2009 via UPS - return receipt. Letters explained the reason for recall and the potential hazard to patients. The letter contains a list of all affected lot numbers as well as a picture which shows that units with a clearly visible blue ring are NOT affected. Distributors are asked to examine inventory and quarantine all affected product and return the response form, even if no recalled product is in inventory. If the product was further distributed, the letter requests that a copy of the letter be sent to all customers and for returns, follow Distributor actions listed in the letter. Retailers and end users/Hospitals/EMT's are asked to perform a count of recalled product, segregate it and return the response form. All customers are asked to call 1-800-634-6003 where they will be assigned a RGA number and given instructions on how to return product. Contact information: 8:00AM - 5:30 PM Central Time Monday through Thursday and 8:00AM to 1:00 PM on Fridays, Central Time.
Quantity
over 200 total lots.
Distribution
Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru, Poland, Russia and Vietnam.
Lot, serial or model codes
Foreign distribution only: Product code: 734-E, lot codes: Japan: 06-29 Argentenia: 03-43 03-44 05-19 05-39 06-08 07-04 07-36.   Product ID # 734MM, lot codes: Russia: 04-25 04-42 05-36 05-38 05-47 06-17 06-30 06-37 06-43 06-49 07-03 07-14
510(k) numbers
["K873286"]
PMA numbers
not on file
Product code
BTM
Firm FEI
1000160240
Firm city
Mcallen
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register