Skip to the content

FDA Medical Device Recalls (openFDA)

Bio-Detek, Inc.: ZOLL Pro-padz Sterile Adult Multi-Function Electrodes with 10-ft (3 m) leadwires, 1 single Part number: 8900-4055-40 Product Usage: These MFE Electrodes are used in connection with ZOLL Defibrillators for Adult patients. These disposable single use MFEs are processed through Gamma Radiation to a SAL of 10-6 per International/ National Standards, EN ISO 11137-2:2012 and ANSI/AAMI/ISO 11137-2:2013, Method 1, for use in a sterile environment by trained medical professionals. The indications for use include defibrillation, cardioversion, external non-invasive pacing and monitoring.

Recall number
Z-0055-2015
Recalling firm
Bio-Detek, Inc.
Product
ZOLL Pro-padz Sterile Adult Multi-Function Electrodes with 10-ft (3 m) leadwires, 1 single Part number: 8900-4055-40 Product Usage: These MFE Electrodes are used in connection with ZOLL Defibrillators for Adult patients. These disposable single use MFEs are processed through Gamma Radiation to a SAL of 10-6 per International/ National Standards, EN ISO 11137-2:2012 and ANSI/AAMI/ISO 11137-2:2013, Method 1, for use in a sterile environment by trained medical professionals. The indications for use include defibrillation, cardioversion, external non-invasive pacing and monitoring.
Status
Terminated
Initiated
2014-09-16
Posted
2014-10-09
Terminated
2016-05-03
Reason
Product sterility is not assured
Root cause
Material/Component Contamination
Action
Bio-Detek notiifed customers of the expanded recall on September 16, 2014 for Lot 1914 and September 22, 2014 for Lots 1614, 1814, 2114 via UPS overnight delivery. Customers were requested to return the suspect product(s) in their possession. A Device Correction Response Return Form was provided to each Customer. For questions call 1-800-348-9011 or 1-978-421-9460.
Quantity
41 singles
Distribution
Worldwide Distribution - US Nationwide and the countries of: Australia, Canada, Belgium, France, and Germany.
Lot, serial or model codes
Lot Numbers: 1614, 1914, 2114
510(k) numbers
["K943698"]
PMA numbers
not on file
Product code
MKJ
Firm FEI
1218058
Firm city
Pawtucket
Firm state
RI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register