Skip to the content

FDA Medical Device Recalls (openFDA)

Ge Healthcare: ApexPro FH Telemetry System: composed of six major components (as follows); Accessories to the patient worn acquisition transceivers, the patient worn data acquisition transceivers, the transceiver access points with antenna, the network infrastructure, A computer platform running the ApexPro Telemetry Application and a computer platform running a central station application (which may be the same computer platform running the ApexPro Telemetry Application). GE Medical Systems information Technologies, 8200 West Tower Avenue, Milwaukee, WI 53223.

Recall number
Z-0054-2007
Recalling firm
Ge Healthcare
Product
ApexPro FH Telemetry System: composed of six major components (as follows); Accessories to the patient worn acquisition transceivers, the patient worn data acquisition transceivers, the transceiver access points with antenna, the network infrastructure, A computer platform running the ApexPro Telemetry Application and a computer platform running a central station application (which may be the same computer platform running the ApexPro Telemetry Application). GE Medical Systems information Technologies, 8200 West Tower Avenue, Milwaukee, WI 53223.
Status
Terminated
Initiated
2006-06-23
Posted
2006-10-21
Terminated
2007-05-25
Reason
System Warning Alarm failure: When a patient being monitored is in a pre-existing condition of continuous MESSAGE or ADVISORY level alarm preceding a SYSTEM WARNING level alarm, the SYSTEM WARNING audible alarm and flashing yellow border around the patient panel at the CIC does not occur.
Root cause
Other
Action
An Urgent Medical Device Correction letter, dated June 23, 2006, was sent to customers notifying them of the issue, provided short term recommendations and provided the long term solution as a software upgrade that will correct this situation. A reply card is asked to be returned to initiate the upgrade process.
Quantity
379
Distribution
Nationwide distribution --- including states of AL, AR, AZ, CA, CO, Washington DC, FL, IL, IN, ME, MI, MN, MO, NC, NE, NJ, NY, OK, PA, SC, TX, VA, WA, WI, WV.
Lot, serial or model codes
Software versions v3.8 and earlier.
510(k) numbers
["K033365"]
PMA numbers
not on file
Product code
MHX
Firm FEI
3005860720
Firm city
Wauwatosa
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register