Skip to the content

FDA Medical Device Recalls (openFDA)

Terumo Cardiovascular Systems Corporation: Sarns Centrifugal System Control module System 9000 The Sarns Centrifugal System is indicated for use in cardiopulmonary bypass procedures only.

Recall number
Z-0053-2013
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Sarns Centrifugal System Control module System 9000 The Sarns Centrifugal System is indicated for use in cardiopulmonary bypass procedures only.
Status
Terminated
Initiated
2012-09-24
Posted
2012-10-11
Terminated
2013-11-13
Reason
Terumo Cardiovascular Systems (Terumo CVS) received one report of loss of power to the control module and subsequently the drive motor for the Sarns centrifugal system during cardiopulmonary bypass.
Root cause
Component design/selection
Action
TERUMO sent an Urgent Medical Device Recall dated April 13, 2012, to all affected consignees. An Addendum was sent on September 25, 2012. All US Consignees will receive a Safety Advisory and an update to the Operator's Manual via Federal Express to inform them of the reported malfunction and potential hazards and to raise awareness of the benefits of using a retrograde valve and the importance of back-up system availability. TERUMO will include a customer response form and follow up with all affected users to obtain a response to the recall. For questions customers should call 1-800-521-2818. For questions regarding this recall call 734-741-6173.
Quantity
126 units
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Bangladesh, Seoul Korea, Chile, Brazil, Venesuela, Thailand, Mexico, India, Taiwan, Canada, Dominican Republic, Guatemala, Ecuador, Germany, Republic of Georgia, Hong Kong, Japan, United Arab Emirates (UAE), Japan, Belgium, Phillipines, Singapore, and South Africa.
Lot, serial or model codes
Catalog number: 6381 and serial numbers: 1003-1006, 1020, 1026, 1045, 2001, 2017,2018,2036,7000-7113, and 7381.
510(k) numbers
["K882758"]
PMA numbers
not on file
Product code
KFM
Firm FEI
1828100
Firm city
Ann Arbor
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register