Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc.: e.cam Emission Imaging Computed Tomography Systems used for all common nuclear medicine procedures; Siemens Medical Systems, Inc., Nuclear Medicine Group, Hoffman Estates, IL 60195-5203

Recall number
Z-0053-04
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
e.cam Emission Imaging Computed Tomography Systems used for all common nuclear medicine procedures; Siemens Medical Systems, Inc., Nuclear Medicine Group, Hoffman Estates, IL 60195-5203
Status
Terminated
Initiated
2003-09-05
Posted
2003-10-24
Terminated
2005-02-24
Reason
The detector's radial brake may not be applied properly, allowing unintended motion which could result in injury to the patient or operator.
Root cause
Other
Action
Siemens sent Important Safety Customer Advisory letters dated 8/27/03 to all e.cam customers on 9/5-12/03. The accounts were informed of the problem with the detector''s radial brake not being applied properly, allowing slow, unintended downward motion of the detector that could result in injury to the patient or operator, and were advised not to leave the patient unattended while on the pallet and not to leave the patient under the detector upon completion of the study. The customers were advised that they would be contacted by their Siemens service provider to schedule a mandatory update to the e.cam to correct the problem. The upgrade will be done free of charge.
Quantity
2,386 units
Distribution
Nationwide, United Arab Emirates, Albania, Argentina, Austria, Australia, Bosnia, Barbados, Bangladesh, Belgium, Bolivia, Brazil, Cameroon, Canada, Chile, China, Colombia, Republic of Congo, Costa Rica, Czech Republic, Denmark, Dominican Republic, El Salvador, Ecuador, Egypt, France, Finland, Germany, Great Britain, Georgia, Greece, Hong Kong, Honduras, Ireland, India, Italy, Iran, Iraq, Jordan, Japan, Korea, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Mexico, Morocco, Macedonia, Madagascar, Myanmar, Mauritius, Malaysia, Nepal, Norway, New Zealand, Oman, Peru, Philippines, Pakistan, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Turkey, Ukraine, Uzbekistan, Zimbabwe, South Africa and the Netherlands.
Lot, serial or model codes
all e.cam models: 4380213, 4380221, 5242826, 5977066, 5977074, 5984005, 5989079, 5989087, 5989095, 5991109, 5991117, 5992099, 7332625, 7332633, 7760809, 7760932, 7761161, 7823920, 7823938, 7823946, 7823953, 7823979; serial numbers 1 through 9091
510(k) numbers
["K952109"]
PMA numbers
not on file
Product code
KPS--
Firm FEI
1000514697
Firm city
Hoffman Estates
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register