K2M, Inc: K2M Special Connector Instruments (Special Connector, Special Adaptor, and Quick Adaptor) Product Usage: Special Connector Instruments include an additional feature on their shafts that enable surgeon users to combine them with non-K2M navigation systems as part of their standard surgical technique.

Recall number
Z-0052-2020
Recalling firm
K2M, Inc
Product
K2M Special Connector Instruments (Special Connector, Special Adaptor, and Quick Adaptor) Product Usage: Special Connector Instruments include an additional feature on their shafts that enable surgeon users to combine them with non-K2M navigation systems as part of their standard surgical technique.
Status
Terminated
Initiated
2016-12-02
Posted
not on file
Terminated
2020-05-22
Reason
Instruments were developed without adequate consideration given to the need for a new regulatory filing. This event was reported to FDA as part of Stryker's retrospective review of K2M product field actions; recall actions were completed in 2017.
Root cause
Nonconforming Material/Component
Action
K2M notified customers with a Recall Notification Letter dated December 2, 2016. The letter instructed customers to locate and segregate recalled product, complete a tracking/verification form, and return the product to K2M. The product was reintroduced to the market after K2M obtained new 510(k) clearance for the product. for question call at 571-919-2000
Quantity
503
Distribution
US Nationwide Distribution in the states of AL CA CO CT FL FL IL IN KS MA MI MN PA TN TX UT VA
Lot, serial or model codes
All product lots in distribution as of December 1, 2016, are in scope.
510(k) numbers
["K171321"]
PMA numbers
not on file
Product code
OLO
Firm FEI
3008599177
Firm city
Leesburg
Firm state
VA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register