Skip to the content

FDA Medical Device Recalls (openFDA)

Bios S.r.l.: THERMI250 Radiofrequency System. Intended to provide heating for the purpose of elevating tissue temperature for selected medical conditions.

Recall number
Z-0052-2018
Recalling firm
Bios S.r.l.
Product
THERMI250 Radiofrequency System. Intended to provide heating for the purpose of elevating tissue temperature for selected medical conditions.
Status
Terminated
Initiated
2017-07-13
Posted
not on file
Terminated
2020-09-15
Reason
In some systems, the unit continues to emit RF energy after the PAUSE button is pressed.
Root cause
Software design
Action
Consignees were sent a recall notification by mail dated 07/13/2017. Customers were notified of the reason for recall and advised that the PAUSE button not be used and the STOP button be used instead to stop the RF emission. Additionally, units are to be connected to the internet so that the BIOS service team can access the units to provide a software update to correct the issue. If a unit cannot be connected to the internet, customers are to contact Thermi to schedule an update.
Quantity
167 units
Distribution
AZ, AR, CA, FL, GA, ID, IL, KS, LA, MA, MI, MS, MO, NV, NJ, NM, NY, OH, OR, PA, SD, TX, WA, WI, and WV Brazil, Canada, Israel, Japan, & Thailand.
Lot, serial or model codes
Serial No. TS250-16G-088-1, TS250-16G-089-1, TS250-16G-090-1, TS250-16G-094-1, TS250-16G-097-1, TS250-16G-098-1, TS250-16G-106-1, TS250-16G-109-2, TS250-16I-116-1, TS250-16I-120-1, TS250-16I-121-1, TS250-16I-122-1, TS250-16I-123-1, TS250-16I-124-1, TS250-16I-127-1, TS250-16I-128-1, TS250-16I-129-1, TS250-16I-130-1, TS250-16I-134-1, TS250-16I-135-1, TS250-16I-136-1, TS250-16I-137-1, TS250-16I-140-1, TS250-16I-142-1, TS250-16I-146-2, TS250-16L-156-1, TS250-16L-158-1, TS250-16L-159-1, TS250-16L-160-1, TS250-16L-164-1, TS250-16L-165-1, TS250-16L-167-1, TS250-16L-168-1, TS250-16L-169-1, TS250-16L-170-1, TS250-16L-177-1, TS250-16L-178-1, TS250-16L-179-1, TS250-16N-190-1, TS250-16N-191-1, TS250-16N-192-1, TS250-16N-194-1, TS250-16N-195-1, TS250-16N-197-1, TS250-16N-198-1, TS250-16N-199-1, TS250-16N-202-1, TS250-16N-203-1, TS250-16N-204-1, TS250-16N-205-1, TS250-16N-207-1, TS250-16N-208-1, TS250-16N-209-1, TS250-16N-210-1, TS250-16N-211-1, TS250-16N-213-1, TS250-16N-214-1, TS250-16N-215-1, TS250-16N-216-1, TS250-16N-217-1, TS250-16N-219-1, TS250-16N-220-1, TS250-16N-223-1, TS250-16N-224-1, TS250-17B-229-1, TS250-17B-230-1, TS250-17B-231-1, TS250-17B-239-1, TS250-17B-243-1, TS250-17B-247-1, TS250-17B-249-1, TS250-17B-250-1, TS250-16L-153-1, TS250-16L-180-2, TS250-16N-212-1, TS250-17B-225-1, TS250-17B-226-1, TS250-17B-228-1, TS250-17B-232-1, TS250-17B-233-1, TS250-17B-234-1, TS250-17B-242-1, TS250-17B-245-1, TS250-16G-093-1, TS250-16G-104-1, TS250-16L-182-2, TS250-17B-235-1, TS250-17B-236-1, TS250-17B-237-1, TS250-17B-238-1, TS250-17B-244-1, TS250-16I-112-2, TS250-16I-113-2, TS250-16I-114-2, TS250-16I-115-2, TS250-16G-100-1, TS250-16G-081-1, TS250-16G-082-1, TS250-16G-083-1, TS250-16G-084-1, TS250-16G-085-1, TS250-16G-086-1, TS250-16G-087-1, TS250-16G-091-1, TS250-16G-092-1, TS250-16G-095-1, TS250-16G-096-1, TS250-16G-099-1, TS250-16G-101-1, TS250-16G-102-1, TS250-16G-103-1, TS250-16G-105-1, TS250-16G-107-1, TS250-16H-110-2, TS250-16I-117-1, TS250-16I-118-1, TS250-16I-119-1, TS250-16I-125-1, TS250-16I-126-1, TS250-16I-131-1, TS250-16I-132-1, TS250-16I-133-1, TS250-16I-138-1, TS250-16I-139-1, TS250-16I-141-1, TS250-16I-143-1, TS250-16I-144-2, TS250-16I-145-2, TS250-16I-147-2, TS250-16L-155-1, TS250-16L-176-1, TS250-16L-183-2, TS250-16N-193-1, TS250-16N-196-1, TS250-16N-200-1, TS250-16N-201-1, TS250-16N-206-1, TS250-16N-218-1, TS250-16N-222-1, TS250-17B-227-1, TS250-17B-240-1, TS250-17B-241-1, TS250-17B-246-1, TS250-17B-248-1, TS250-17B-251-1, TS250-17B-252-1, TS250-17B-253-1, TS250-17B-254-1, TS250-17B-255-1, TS250-17B-256-1, TS250-17B-257-1, TS250-17B-258-1, TS250-17B-259-1, TS250-17B-260-1, TS250-16H-108-2, TS250-16L-148-1, TS250-16L-149-1, TS250-16L-150-1, TS250-16L-151-1, TS250-16L-152-1, TS250-16L-157-1, S250-16L-181-2, TS250-16L-184-1, TS250-16L-185-1, TS250-16L-186-1, TS250-16L-187-1, and TS250-16L-188-1
510(k) numbers
["K151296"]
PMA numbers
not on file
Product code
PBX
Firm FEI
3008108972
Firm city
Vimodrone
Firm state
not on file
Firm country
Italy

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register