Skip to the content

FDA Medical Device Recalls (openFDA)

St Jude Medical: Pulse Generator

Recall number
Z-0052-04
Recalling firm
St Jude Medical
Product
Pulse Generator
Status
Terminated
Initiated
2003-07-31
Posted
2003-10-24
Terminated
2004-03-26
Reason
Under certain circumstances, pacemaker has a potential to deliver a short coupled pacing interval of approx. 300 msec (200ppm). It is also possible that a patient may experience (up to max of 12) shorter than anticipated pacing intervals.
Root cause
Other
Action
Sales Representatives will personally contact the physicians, explain the situation, leave a technical memo and a list of implanted patients for the doctor to follow-up with. The ''Dear Doctor'' letter to accompany the representatives is dated 7/31/2003. Instruments not implanted will be returned for reprocessing.
Quantity
775
Distribution
Nationwide and to four VA Medical Centers in IA and MO.
Lot, serial or model codes
Integrity ADx DR Model 5360 & 5366 Identity ADx DR model 5286, 5380, 5386 &5480 Verity ADx DR 5256, 5356 & 5456
510(k) numbers
not on file
PMA numbers
["P880086"]
Product code
DXY
Firm FEI
2017865
Firm city
Sylmar
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register