FDA Medical Device Recalls (openFDA)
Abbott Laboratories Medisense Products: Xtra Classic Glucose Monitoring System; Recalling Firm/Manufacturer: Abbott Diabetes Care, Inc., 1360 South Loop Road, Alameda, CA 94502
- Recall number
- Z-0051-06
- Recalling firm
- Abbott Laboratories Medisense Products
- Product
- Xtra Classic Glucose Monitoring System; Recalling Firm/Manufacturer: Abbott Diabetes Care, Inc., 1360 South Loop Road, Alameda, CA 94502
- Status
- Terminated
- Initiated
- 2005-06-07
- Posted
- 2005-10-21
- Terminated
- 2006-05-17
- Reason
- User may inadvertently change the Units of Measure from mg/dL to mmol/L and the blood glucose results could be misinterpreted. This may lead to under treatment and potential for hyperglycemia.
- Root cause
- Other
- Action
- The firm has undertaken a field correction to notify their customers of the potential for the units of measure to be changed. The firm plans to (1) include a flyer in all Precision Xtra and FreeStyle meter kits to reinforce current labeling regarding units of measure and update the Frequently Asked Questions section of the website to reinforce the units of measure change, (2) Lock out the units of measure in new production meters to eliminate potential to change the units of measure settings, update the meter kits with a flyer to override the section of the manual indicating that the units of measure are changeable, and change the product insert in glucose test strips to indicate which units of measure are appropriate for the user and (3) update the labeling to indicate that the units of measure are not user configurable to eliminate the need for the flier.
- Quantity
- not on file
- Distribution
- Approximately 10,005,206 meters have been distributed worldwide since 1999, with 5,885,332 meters distributed in the US, and 4,517,211 in use.
- Lot, serial or model codes
- All lot numbers are included in this recall.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- NBW
- Firm FEI
- 3002803444
- Firm city
- Alameda
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28