Skip to the content

FDA Medical Device Recalls (openFDA)

Nucletron BV: Oncentra Brachy 4.5 radiation therapy software The firm name on the label is Nucletron B.V. Oncentra is a radiation therapy planning software designed to analyze and plan radiation treatments in three dimensions for the purpose of treating patients with cancer. The treatment plans provide estimates of dose distributions expected during the proposed treatment and may be used to administer treatments after review and approval by qualified medical personnel.

Recall number
Z-0050-2018
Recalling firm
Nucletron BV
Product
Oncentra Brachy 4.5 radiation therapy software The firm name on the label is Nucletron B.V. Oncentra is a radiation therapy planning software designed to analyze and plan radiation treatments in three dimensions for the purpose of treating patients with cancer. The treatment plans provide estimates of dose distributions expected during the proposed treatment and may be used to administer treatments after review and approval by qualified medical personnel.
Status
Terminated
Initiated
2017-08-10
Posted
not on file
Terminated
2020-11-20
Reason
Incorrect source step size may occur in the software plans
Root cause
Software design
Action
Elekta sent an Field Safety Notice notification dated August 2017, and was emailed to customers on August 10, 2017, informed users of the issue and workarounds that can be used to avoid the issue.The user is informed the issue will be resolved in a fix which will be released by the end of February 2018. Once released, the firm will proactively reach out to all affected customers to arrange a time to install the fix. An acknowledgement form was enclosed which the user was to complete to indicate they have read and understood the notice and accept implementation of any given recommendations. For further questions, please call (770) 670-2359.
Quantity
89 customer sites have the affected software
Distribution
Worldwide Distribution - USA (nationwide) to medical facilities, and to the countries of : Canada, Argentina, Australia, Austria Bangladesh, Belarus, Belgium, Brazil, Brunei Darussalam, Bulgaria, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Kosovo, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Pakistan, Peru, Philippines, Poland, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, Ukraine, United Arab Emirates, and Viet Nam. There was no military/government distribution.
Lot, serial or model codes
Versions 4.5, 4.5.1, and 4.5.2.
510(k) numbers
["K132816"]
PMA numbers
not on file
Product code
MUJ
Firm FEI
3002807741
Firm city
Veenendaal
Firm state
not on file
Firm country
Netherlands

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register