Stryker Howmedica Osteonics Corp.: Scorpio/ NRG 4in1 Cutting Block The Scorpio total knee system is indicated for primary Total knee Arthoplasty (TKA) procedures. The Scorpio Femoral Component is implanted on femur after making bone cuts to accommodate the shape of the implant. The Scorpio instrumentation includes a 4 in 1 Cutting block 98010-00xx) which serves as a guide to allow the surgeon to make the bone e cuts necessary for seating the femoral component (implant) in place. The Scorpio Cutting blocks (8010-00xx) are designed to align with two previously located holes and for seating on the distal surg ace of the femur, and provide guided slots for the sawblades to make bone cuts.

Recall number
Z-0050-2011
Recalling firm
Stryker Howmedica Osteonics Corp.
Product
Scorpio/ NRG 4in1 Cutting Block The Scorpio total knee system is indicated for primary Total knee Arthoplasty (TKA) procedures. The Scorpio Femoral Component is implanted on femur after making bone cuts to accommodate the shape of the implant. The Scorpio instrumentation includes a 4 in 1 Cutting block 98010-00xx) which serves as a guide to allow the surgeon to make the bone e cuts necessary for seating the femoral component (implant) in place. The Scorpio Cutting blocks (8010-00xx) are designed to align with two previously located holes and for seating on the distal surg ace of the femur, and provide guided slots for the sawblades to make bone cuts.
Status
Terminated
Initiated
2010-08-23
Posted
2010-10-13
Terminated
2012-08-30
Reason
Stryker s has received reports in which the Cutting Blocks have fractured on impaction during surgery.
Root cause
Device Design
Action
Stryker Orthopaedics sent an "URGENT PRODUCT RECALL" letter dated August 30, 2010, to all consignees/customers (branches/agencies, hospital risk management, hospital Chief of Orthopaedics, surgery) including foreign countries. The letter described the product, problem, and action to be taken by customers. The customers were instructed to complete and return the attached PRODUCT RECALL ACKNOWLEDGMENT FORM via fax to 201-831-6069. If you have any questions, feel free to contact the Manager, Divisional Regulatory Compliance (201) 972-2100.
Quantity
839 (total)
Distribution
Worldwide distribution: USA and countries including: Asia, Canada, Europe, Middle East and Africa
Lot, serial or model codes
8010-0004
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
3003070421
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register