Becton Dickinson & Company: CareFusion MaxPlus Clear Needleless Connector. Intended for Positive Displacement Connector for needleless access to the IV line and/or IV catheter during IV therapy Cat. No. MP1000-C
- Recall number
- Z-0049-2020
- Recalling firm
- Becton Dickinson & Company
- Product
- CareFusion MaxPlus Clear Needleless Connector. Intended for Positive Displacement Connector for needleless access to the IV line and/or IV catheter during IV therapy Cat. No. MP1000-C
- Status
- Terminated
- Initiated
- 2019-05-02
- Posted
- not on file
- Terminated
- 2021-11-30
- Reason
- After disconnection of a luer, the valve of the connector remains recessed; not fully closing, either temporarily or permanently. The risks of blood loss and air embolism exist if the valve remains permanently open and goes undetected, especially in the neonate population.
- Root cause
- Under Investigation by firm
- Action
- BD MDS (Medication Delivery Solutions) initiated communication to US on May 2, 2019 via Fed'X. Letter states product defect, health risk and action to take:1. Immediately review for the specific Catalogue (Ref) and lot numbers . Destroy all product subject to the recall following your institutions process for destruction. 2. Share this recall notification with all users of the product within your facility to ensure that they are also aware of this recall. 3. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification and process your product replacement. Regional Customer Quality 888-237-2762 OPT 3, OPT 2 Monday Friday between the hours of 8:00am and 5:00pm (CST)
- Quantity
- 115,200
- Distribution
- Nationwide Foreign: Canada
- Lot, serial or model codes
- Lot Number: 18085990; Exp. Date: 15-Aug-2023 UDI: 10885403223228
- 510(k) numbers
- ["K083472"]
- PMA numbers
- not on file
- Product code
- FPA
- Firm FEI
- 2243072
- Firm city
- Franklin Lakes
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28