Boston Scientific CRM Corp: Guidant ASSURE, Model B301, Implantable Cardioverter Defibrillator, ICD. (This model is not available in the USA) Guidant Corporation Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798, USA.

Recall number
Z-0049-2008
Recalling firm
Boston Scientific CRM Corp
Product
Guidant ASSURE, Model B301, Implantable Cardioverter Defibrillator, ICD. (This model is not available in the USA) Guidant Corporation Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798, USA.
Status
Terminated
Initiated
2007-03-16
Posted
2007-11-24
Terminated
2007-11-23
Reason
Atypical Charge Time Behavior: 1)-End of Life (EOL) or ERI is displayed during mid-life (typically 24-48 months), even though battery capacity remains available. 2)- Extended ERI charge time limits: Charge times during mid-life may remain below a normal extension of the ERI charge time . ****** Prophylactic replacement of these devices prior to ERI is not recommended. As stated in the Product
Root cause
Device Design
Action
A Product Update titled "ERI Charge Time Limit Extended During Mid-Life and Mid-Life Display of Replacement Indicators", dated 03/10/2007 was posted to Company's website on 03/13/2007. A CRM Product News letter was sent to physicians beginning 03/16/2007. The CRM Product News Letter gave an overview of the content of the Product Update and stated the web location for finding the complete Product Update. The Product Update identified affected devices, gave an overview of normal charge time behavior at mid-life for specific ICD and CRT-D devices. The letter also described the atypical charge time behavior, projected rate of mid-life display of replacement indicators and Patient Management Considerations. ****** Prophylactic replacement of these devices prior to ERI is not recommended. As stated in the Product Update, "Devices that have triggered charge time-based ERI or EOL during mid-life have several months, and in most cases more than one year of remaining battery voltage and capacity, which allows the devices in this pattern to continue to provide brady and LV pacing and maximum energy shocks. However, if ERI or EOL is triggered, device replacement should be scheduled."*****Recalled units only cover devices implanted prior to July 2005****
Quantity
180-Recalled units only cover devices implanted prior to July 2005
Distribution
Worldwide
Lot, serial or model codes
Model B301: B301/100002, B301/100003, B301/100004, B301/100006, B301/100007, B301/100008, B301/100010, B301/100012, B301/100013, B301/100014, B301/100016, B301/100019, B301/100020, B301/100021, B301/100022, B301/100024, B301/100025, B301/100027, B301/100030, B301/100031, B301/100032, B301/100034, B301/100037, B301/100039, B301/100043, B301/100044, B301/100045, B301/100046, B301/100048, B301/100050, B301/100051, B301/100053, B301/100054, B301/100056, B301/100062, B301/100063, B301/100064, B301/100067, B301/100069, B301/100070, B301/100073, B301/100074, B301/100075, B301/100076, B301/100079, B301/100080, B301/100081, B301/100082, B301/100083, B301/100084, B301/100087, B301/100088, B301/100089, B301/100090, B301/100091, B301/100092, B301/100094, B301/100096, B301/100099, B301/100100, B301/100101, B301/100102, B301/100103, B301/100104, B301/100105, B301/100106, B301/100108, B301/100111, B301/100112, B301/100113, B301/100114, B301/100115, B301/100116, B301/100117, B301/100118, B301/100120, B301/100122, B301/100125, B301/100126, B301/100128, B301/100129, B301/100130, B301/100131, B301/100132, B301/100134, B301/100135, B301/100136, B301/100137, B301/100139, B301/100140, B301/100141, B301/100142, B301/100143, B301/100144, B301/100145, B301/100146, B301/100147, B301/100148, B301/100149, B301/100153, B301/100154, B301/100155, B301/100158, B301/100159, B301/100161, B301/100162, B301/100164, B301/100165, B301/100166, B301/100167, B301/100169, B301/100170, B301/100172, B301/100173, B301/100174, B301/100176, B301/100180, B301/100181, B301/100182, B301/100183, B301/100185, B301/100186, B301/100187, B301/100188, B301/100189, B301/100191, B301/100193, B301/100194, B301/100195, B301/100196, B301/100197, B301/100200, B301/100203, B301/100204, B301/100205, B301/100206, B301/100208, B301/100209, B301/100210, B301/100211, B301/100213, B301/100215, B301/100216, B301/100218, B301/100221, B301/100222, B301/100223, B301/100224, B301/100226, B301/100227, B301/100230, B301/100234, B301/100235, B301/100236, B301/100238, B301/100239, B301/100242, B301/100243, B301/100244, B301/100245, B301/100247, B301/100249, B301/100253, B301/100255, B301/100258, B301/100260, B301/100264, B301/100271, B301/100272, B301/100273, B301/100279, B301/100280, B301/100285, B301/100288, B301/100290, B301/100293, B301/100298, B301/100315, B301/100317, B301/100318
510(k) numbers
not on file
PMA numbers
not on file
Product code
LWP
Firm FEI
2124215
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register