Siemens Medical Solutions USA, Inc.: Torque wrench used on the bolts for the Symbia T2 model 10275008, Symbia Intevo Excel model 10764801, Symbia Intevo 16 model 10764804, Symbia Evo model 10910500, and Symbia Evo Excel model 10910501 devices.
- Recall number
- Z-0047-2018
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Torque wrench used on the bolts for the Symbia T2 model 10275008, Symbia Intevo Excel model 10764801, Symbia Intevo 16 model 10764804, Symbia Evo model 10910500, and Symbia Evo Excel model 10910501 devices.
- Status
- Terminated
- Initiated
- 2017-09-18
- Posted
- not on file
- Terminated
- 2017-12-11
- Reason
- The torque wrench used to tighten system bolts during installation was found to be out of tolerance
- Root cause
- Equipment maintenance
- Action
- The recall was initiated via phone and visit beginning 9/18/2017.
- Quantity
- 16 devices
- Distribution
- Distribution was made to medical facilities located in CA, FL, GA, LA, NY, PA, and TN. There was government distribution but no foreign/military distribution.
- Lot, serial or model codes
- Symbia T2 serial number 2001; Symbia Intevo Excel serial numbers 2112 and 2140; Symbia Intevo 16 serial numbers 2100 and 2116; Symbia Evo serial numbers 1049, 1054, 1071, 1075, 1104, and 1128; and Symbia Evo Excel serial numbers 1109, 1111, 1131, 1134, and 1158.
- 510(k) numbers
- ["K162337"]
- PMA numbers
- not on file
- Product code
- KPS
- Firm FEI
- 1000514697
- Firm city
- Hoffman Estates
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28