FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc.: Torque wrench used on the bolts for the Symbia T2 model 10275008, Symbia Intevo Excel model 10764801, Symbia Intevo 16 model 10764804, Symbia Evo model 10910500, and Symbia Evo Excel model 10910501 devices.

Recall number
Z-0047-2018
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Torque wrench used on the bolts for the Symbia T2 model 10275008, Symbia Intevo Excel model 10764801, Symbia Intevo 16 model 10764804, Symbia Evo model 10910500, and Symbia Evo Excel model 10910501 devices.
Status
Terminated
Initiated
2017-09-18
Posted
not on file
Terminated
2017-12-11
Reason
The torque wrench used to tighten system bolts during installation was found to be out of tolerance
Root cause
Equipment maintenance
Action
The recall was initiated via phone and visit beginning 9/18/2017.
Quantity
16 devices
Distribution
Distribution was made to medical facilities located in CA, FL, GA, LA, NY, PA, and TN. There was government distribution but no foreign/military distribution.
Lot, serial or model codes
Symbia T2 serial number 2001; Symbia Intevo Excel serial numbers 2112 and 2140; Symbia Intevo 16 serial numbers 2100 and 2116; Symbia Evo serial numbers 1049, 1054, 1071, 1075, 1104, and 1128; and Symbia Evo Excel serial numbers 1109, 1111, 1131, 1134, and 1158.
510(k) numbers
["K162337"]
PMA numbers
not on file
Product code
KPS
Firm FEI
1000514697
Firm city
Hoffman Estates
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register