Skip to the content

FDA Medical Device Recalls (openFDA)

Interventional Spine Inc: PERPOS Driver; Catalog # 6112 Interventional Spine, Inc. A cannulated, driver to be used over a K-wire to engage the head of the 4.5mm BONE-LOK PLS Implant (LSW-45-3040) and drive it into the bone. The PERPOS Driver (61 12) is sold as either a component of the Single Use PERPOS PLS System (9045-01, 9045-02), a component of the Reusable BONE-LOK PLS System (9039-01, 9039-02), or as a stand-alone tool.

Recall number
Z-0047-2012
Recalling firm
Interventional Spine Inc
Product
PERPOS Driver; Catalog # 6112 Interventional Spine, Inc. A cannulated, driver to be used over a K-wire to engage the head of the 4.5mm BONE-LOK PLS Implant (LSW-45-3040) and drive it into the bone. The PERPOS Driver (61 12) is sold as either a component of the Single Use PERPOS PLS System (9045-01, 9045-02), a component of the Reusable BONE-LOK PLS System (9039-01, 9039-02), or as a stand-alone tool.
Status
Terminated
Initiated
2011-02-03
Posted
2011-10-13
Terminated
2013-02-05
Reason
This recall has been initiated due to drivers that may contain a burr in the distal tip preventing full seating of the driver onto the head of the facet screw, resulting in difficulty screwing the facet screw into the bone. Use of this product may result in a delay of the surgical procedure
Root cause
Nonconforming Material/Component
Action
Interventional Spine, Inc. sent an "Urgent Recall" letter dated March 15, 2011 to all affected customers. The letter included; affected product, reason for recall and recommendations. Customers were asked to immediately examine inventory, quarantine affected product and to notify International Spine, Inc if any product remain. For information on this recall contact the firm at (949) 472-0006.
Quantity
12 units
Distribution
Word wide Distribution: (USA) Nationwide including the states of CO, CT, FL, KY, LA, MA, NC, NY, OK, OR, TN, TX, VA, and WI; and the countries of Italy, Spain, Turkey, and the U.K.
Lot, serial or model codes
DRIVER CATALOG NO. 6112; 050507A, 061509A, 071808A, 112507A, 042709E
510(k) numbers
not on file
PMA numbers
not on file
Product code
HXX
Firm FEI
3003523239
Firm city
Irvine
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register