FDA Medical Device Recalls (openFDA)
Beckman Coulter Inc.: Access Folate Calibrators, Part Number: A14207. The Access Folate assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of folic acid levels in human serum, plasma (heparin) or red blood cells using the Access Immunoassay Systems. The serum folate level is an indicator of recent folate intake. A low RBC folate value can indicate a prolonged folate deficiency. The Access Folate Calibrators are intended to calibrate the Access Folate assay using the Access Immunoassay Systems.
- Recall number
- Z-0046-2013
- Recalling firm
- Beckman Coulter Inc.
- Product
- Access Folate Calibrators, Part Number: A14207. The Access Folate assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of folic acid levels in human serum, plasma (heparin) or red blood cells using the Access Immunoassay Systems. The serum folate level is an indicator of recent folate intake. A low RBC folate value can indicate a prolonged folate deficiency. The Access Folate Calibrators are intended to calibrate the Access Folate assay using the Access Immunoassay Systems.
- Status
- Terminated
- Initiated
- 2012-07-24
- Posted
- 2012-10-10
- Terminated
- 2013-10-17
- Reason
- Beckman Coulter is recalling the Access Folate Calibrators because it may produce falsely elevated results that could mask a folate deficiency.
- Root cause
- Nonconforming Material/Component
- Action
- Beckman Coulter sent an "URGENT PRODUCT CORRECTION" letter dated July 24, 2012 to all affected customers. The letter identified the product, problem, and the actions to be taken by the customers. Customers were instructed to complete and return the enclosed Response Forms within 10 days. Customers with questions were instructed to call (800) 854-3633 in the US and Canada. Outside of the US and Canada, customers were instructed to contact their local Beckman Coulter representative.
- Quantity
- 4,635 units total (3,214 units in US)
- Distribution
- Worldwide Distribution-USA (nationwide) and the countries of Argentina, Austria, Australia, Bangladesh, Belgium, Botswana, Brazil, Canada, China, Colombia, Czech Republic, Croatia, Ecuador, Ethiopia, Finland, France, Georgia, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Libyan Arab Jamahiriya, Malaysia, Mexico, Morocco, Namibia, Netherlands, New Zealand, Oman, Pakistan, Panama, Philippines, Poland, Puerto Rico, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Switzerland, Sweden, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, Venezuela, Viet Nam, and Zimbabwe.
- Lot, serial or model codes
- Lot# 112788, 115741, 118917
- 510(k) numbers
- ["K060774"]
- PMA numbers
- not on file
- Product code
- CGN
- Firm FEI
- 2050012
- Firm city
- Brea
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28