Skip to the content

FDA Medical Device Recalls (openFDA)

LDR Spine USA, Inc.: ROI-A Anterior Delivery Device. Orthopedic manual surgical instrument. Model number: SI-ROIA-0023 Auxiliary instrument used in the ROI-A-ALIR cage system: Spinal intervertbral body fusion device.

Recall number
Z-0045-2017
Recalling firm
LDR Spine USA, Inc.
Product
ROI-A Anterior Delivery Device. Orthopedic manual surgical instrument. Model number: SI-ROIA-0023 Auxiliary instrument used in the ROI-A-ALIR cage system: Spinal intervertbral body fusion device.
Status
Terminated
Initiated
2016-08-29
Posted
2016-10-05
Terminated
2017-01-13
Reason
The ROI-A Anterior Delivery Device T-Handle, part number SI-ROIA-0023 has experienced binding of the rotation of the t-handle. There have been no reports of patient injury
Root cause
Device Design
Action
The firm, LDR, contacted affected consignees via phone and then sent a follow up email dated 9/8/16. The phone script and email described the product, problem and actions to be taken. The consignees were instructed to promptly return the instrument using the RMA#1180. The firm will perform a rework that prevents the binding to occur. If you have any question, contact the Quality Engineering Manager by email: ron.musselman@zimmerbiomet.com, phone: 512-344-3436 or Field Inventory Manager - US Supply Chain or phone: 512-344-3436.
Quantity
24 units
Distribution
US Distribution to states of: GA, OR, FL, KY, NC, AR and PA.
Lot, serial or model codes
Lot # 2295101A, 2295101A-R, 2296101A-R
510(k) numbers
["K082262","K090507","K113027","K153495"]
PMA numbers
not on file
Product code
LXH
Firm FEI
3004903783
Firm city
Austin
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register