Skip to the content

FDA Medical Device Recalls (openFDA)

Bard Peripheral Vascular Inc: BARD MAX-CORE Disposable Core Biopsy Instrument, Product Codes - MC1410, 14 g x 10 cm; MC1616, 16 g x 16 cm; MC1816, 18 g x 16 cm; MC1820, 18 g x 20 cm; MC1825, 18 g x 25 cm. The core needle biopsy device is intended for use in obtaining biopsies from soft tissues.

Recall number
Z-0044-2018
Recalling firm
Bard Peripheral Vascular Inc
Product
BARD MAX-CORE Disposable Core Biopsy Instrument, Product Codes - MC1410, 14 g x 10 cm; MC1616, 16 g x 16 cm; MC1816, 18 g x 16 cm; MC1820, 18 g x 20 cm; MC1825, 18 g x 25 cm. The core needle biopsy device is intended for use in obtaining biopsies from soft tissues.
Status
Terminated
Initiated
2017-09-22
Posted
not on file
Terminated
2018-09-27
Reason
During the review of May 2017 complaint data, a significant increase in MAX-CORE Disposable Core Biopsy Instrument complaints was observed for Failure to Prime, Failure to Fire, Failure to Obtain Sample, and Self-Activation.
Root cause
Device Design
Action
BPV sent a customer recall letter on September 22, 2017 to each of the US consignees via FedEx with proof of delivery notification. BPV will also initiate a recall for product outside the US in accordance with regulations of each country in which the product was distributed.
Quantity
110,550 units
Distribution
US, Australia, Canada, Europe, Middle East, Africa, Latin America, Asia Pacific
Lot, serial or model codes
Lot Numbers: REBP0019, REBN0342, REBN2123, REBP1199, REBP1419, REBP1807, REBQ0084, REBQ0343, REBQ1012, REBP1420, REBP1809, REBP0869, REBP1266, REBP1267, REBP1421, REBP1422, REBP1810, REBQ0087, REBQ0088, REBQ0347, REBQ0811, REBQ1014, REBP0158, REBQ1904, REBQ1898, REBQ1978, REBR0468, REBQ2296, REBR0474
510(k) numbers
["K133948"]
PMA numbers
not on file
Product code
KNW
Firm FEI
2020394
Firm city
Tempe
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register