Skip to the content

FDA Medical Device Recalls (openFDA)

Encore Medical, Lp: EMPOWR PS KNEE(TM) e+, REF 343-19-707, SIZE 7, 19mm, QTY 01, GREY, STERILE Product Usage The EMPOWR PS Inserts are manufactured from ultra high molecular weight polyethylene (UHMWPE) infused with pure liquid pharmaceutical grade alpha-tocopheral and then cross-linked to conform to ASTM F2695. This insert is posterior stabilized and is intended to more closely match the kinematics of the knee, allowing some rotation along the medial condyle and increased congruency along the lateral condyle.

Recall number
Z-0042-2018
Recalling firm
Encore Medical, Lp
Product
EMPOWR PS KNEE(TM) e+, REF 343-19-707, SIZE 7, 19mm, QTY 01, GREY, STERILE Product Usage The EMPOWR PS Inserts are manufactured from ultra high molecular weight polyethylene (UHMWPE) infused with pure liquid pharmaceutical grade alpha-tocopheral and then cross-linked to conform to ASTM F2695. This insert is posterior stabilized and is intended to more closely match the kinematics of the knee, allowing some rotation along the medial condyle and increased congruency along the lateral condyle.
Status
Terminated
Initiated
2017-09-15
Posted
not on file
Terminated
2023-08-29
Reason
It was reported that an implant labeled as 343-13-708 was packaged with implant 343-19-707.
Root cause
Packaging process control
Action
The firm initiated the recall by email on 09/05/2017. The letter identified the affected product, problem and actions to be taken. The consignee was directed to notify anyone to whom the product was transferred and to return any in-stock product.
Quantity
46 units
Distribution
US Nationwide Distribution
Lot, serial or model codes
Lot Number 055U1003
510(k) numbers
["K160342"]
PMA numbers
not on file
Product code
JWH
Firm FEI
1000116912
Firm city
Austin
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register