Skip to the content

FDA Medical Device Recalls (openFDA)

Smith & Nephew, Inc.: RENASYS EZ/ RENASYS EZ Plus 800 mL canister (w/o CLP) with Solidifier Product Usage: RENASYS EZ and RENASYS EZ Plus Canisters are used for the collection of exudate removed from wounds being treated by negative pressure wound therapy (NPWT).

Recall number
Z-0041-2015
Recalling firm
Smith & Nephew, Inc.
Product
RENASYS EZ/ RENASYS EZ Plus 800 mL canister (w/o CLP) with Solidifier Product Usage: RENASYS EZ and RENASYS EZ Plus Canisters are used for the collection of exudate removed from wounds being treated by negative pressure wound therapy (NPWT).
Status
Terminated
Initiated
2014-07-18
Posted
2014-10-08
Terminated
2015-08-05
Reason
RENASYS EZ Canisters exhibiting visible evidence of deformation of the inlet port on the canister lid.
Root cause
Device Design
Action
Smith & Nephew sent an Urgent Medical Device Correction Notice dated July 15, 2014 to all US customers via Federal Express delivery. The letter identified the affected product, problem and actions to be taken. The firm instructed customers to discard the affected product and replace with a new one from an unaffected lot. For questions contact 727-399-3785.
Quantity
45,466
Distribution
Worldwide Distribution - US Nationwide in the states of FL, NC, CA, MO, MA, TX, NY, OH, CO, VA, IL, NJ, TN, NV, MN including Puerto Rico and countries of Canada, Mexico, Dubai, Argentina, Australia, Colombia, Germany, Hong Kong, Jordon, Malaysia, Thailand, Tunisia, New Zealand, South Africa, Shanghai, Brazil, Saudi Arabia and Lebanon.
Lot, serial or model codes
Item 66800912 Lot # M400071
510(k) numbers
["K082426","K102001"]
PMA numbers
not on file
Product code
OMP
Firm FEI
3006760724
Firm city
Saint Petersburg
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register