AZIYO BIOLOGICS: CanGaroo Envelopes labeled as the following: 1. CanGaroo Envelope (Large) (single pack); 2. CanGaroo Envelope (Medium) (single pack); 3. CanGaroo Envelope (Extra Large) (single pack); 4. CanGaroo Envelope (Extra Extra Large) (single pack); 5. CanGaroo Envelope (Large) (5 pack); 6. CanGaroo Envelope (Medium) (5 pack); 7. CanGaroo Envelope (Extra Large) (5 pack); 8. CanGaroo Envelope (Extra Extra Large) (5 pack) Product Usage: The CanGaroo Envelope is intended to securely hold a cardiac implantable electronic device or an implantable neurostimulator to create a stable environment when implanted in the body. The cardiac implantable electronic devices that may be used with the CanGaroo Envelope include pacemaker pulse generators, defibrillators, or other cardiac implantable electronic devices. The implantable neurostimulator devices that may be used with the CanGaroo Envelope include vagus nerve stimulators, spinal cord neuromodulators, deep brain stimulators and sacral nerve stimulators.
- Recall number
- Z-0040-2020
- Recalling firm
- AZIYO BIOLOGICS
- Product
- CanGaroo Envelopes labeled as the following: 1. CanGaroo Envelope (Large) (single pack); 2. CanGaroo Envelope (Medium) (single pack); 3. CanGaroo Envelope (Extra Large) (single pack); 4. CanGaroo Envelope (Extra Extra Large) (single pack); 5. CanGaroo Envelope (Large) (5 pack); 6. CanGaroo Envelope (Medium) (5 pack); 7. CanGaroo Envelope (Extra Large) (5 pack); 8. CanGaroo Envelope (Extra Extra Large) (5 pack) Product Usage: The CanGaroo Envelope is intended to securely hold a cardiac implantable electronic device or an implantable neurostimulator to create a stable environment when implanted in the body. The cardiac implantable electronic devices that may be used with the CanGaroo Envelope include pacemaker pulse generators, defibrillators, or other cardiac implantable electronic devices. The implantable neurostimulator devices that may be used with the CanGaroo Envelope include vagus nerve stimulators, spinal cord neuromodulators, deep brain stimulators and sacral nerve stimulators.
- Status
- Terminated
- Initiated
- 2019-08-21
- Posted
- not on file
- Terminated
- 2020-09-17
- Reason
- Possible suture brittleness or breakage along the Envelope seam before the expiration date of the device.
- Root cause
- Under Investigation by firm
- Action
- Aziyo Biologics notified customers on about 08/26/2019 via letter. The letter informed customers of potential suture brittleness or breakage along the envelope seam prior to the expiration of the device. Customers were instructed to immediately discontinue use of affected lots and return them to Aziyo for replacement at no cost. Customers are directed to contact Customer Service at 877-651-2628 to initiate the return, using RGA # 19-1994R. Additionally, customers are instructed to acknowledge of receipt of the recall notice by email (customerservice@aziyo.com), fax (678-680-5486), or telephone (877-651-2628), and return of the Acknowledgement and Receipt Form attached to the recall notification. An updated consignee letter was sent on about 09/04/2019. The letter contained the same information and instructions but notified customers that this action was a removal and not a market withdrawal.
- Quantity
- 2693 units
- Distribution
- US Nationwide distribution including the states of AL, AR, AZ, CA, FL, GA, HI, IL, IN, KS, KY, LA, LA, MI, MN, MO, MS, NC, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA, WI.
- Lot, serial or model codes
- 1. Catalog Number CMCV-009-LRG (UDI Number 000859389005065) Lot Numbers: M18B1053, M18D1080, M18D1081, M18D1091, M18D1095, M19B1077, M19C1083, M19D1144, M19D1148; 2. Catalog Number CMCV-009-MED (UDI Number 000859389005034) Lot Numbers: M18C1071, M18D1088; 3. Catalog Number CMCV-009-XLG (UDI Number 000859389005072) Lot Numbers: M18B1051, M18B1052, M18B1054, M18C1068, M18C1069, M19D1160, M19D1163; 4. Catalog Number CMCV-009-XXL (UDI Number 000859389005119) Lot Numbers: M18D1094, M19C1101, M19C1123; 5. Catalog Number CMCV-010-LRG (UDI Number 010859389005062) Lot Numbers: M18A1007, M18M1303, M18N1330, M19B1071, M19C1104, M19C1119; 6. Catalog Number CMCV-010-MED (UDI Number 010859389005031) Lot Number M18A1015; 7. Catalog Number CMCV-010-XLG (UDI Number 010859389005079) Lot Numbers: M18F1160, M18F1161; 8. Catalog Number CMCV-010-XXL (UDI Number 010859389005116) Lot Number M18F1159
- 510(k) numbers
- ["K140306","K173242","K182255"]
- PMA numbers
- not on file
- Product code
- FTM
- Firm FEI
- 3005619880
- Firm city
- Roswell
- Firm state
- GA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28