Skip to the content

FDA Medical Device Recalls (openFDA)

Genesys Orthopedics Systems, LLC: Genesys spine, Binary Screwdriver, GCP221, component of the Binary Anterior Cervical Plating System The Genesys Spine Binary Anterior Cervical Plate System is intended for anterior screw fixation to the cervical spine. It is to be used in skeletally mature patients as an adjunct to fusion of the cervical spine (C2 to C7). The system is indicated for use in the temporary stabilization of the anterior spine during the development of cervical spinal fusion in patients with degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), spondylolisthesis, trauma (i.e. fractures or dislocations), tumors, deformity (defined as kyphosis, lordosis, or scoliosis),pseudoarthrosis, failed previous fusion, and/or spinal stenosis.

Recall number
Z-0040-2018
Recalling firm
Genesys Orthopedics Systems, LLC
Product
Genesys spine, Binary Screwdriver, GCP221, component of the Binary Anterior Cervical Plating System The Genesys Spine Binary Anterior Cervical Plate System is intended for anterior screw fixation to the cervical spine. It is to be used in skeletally mature patients as an adjunct to fusion of the cervical spine (C2 to C7). The system is indicated for use in the temporary stabilization of the anterior spine during the development of cervical spinal fusion in patients with degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), spondylolisthesis, trauma (i.e. fractures or dislocations), tumors, deformity (defined as kyphosis, lordosis, or scoliosis),pseudoarthrosis, failed previous fusion, and/or spinal stenosis.
Status
Terminated
Initiated
2015-11-05
Posted
not on file
Terminated
2018-07-19
Reason
Unanticipated technique for a cervical screw insertion caused the tip of the threaded inner shaft to break off.
Root cause
Component design/selection
Action
Genesys Spine initiated their field action on November 15, 2015, by letter. The letter explained how and why the instrument failed and provided updated instructions on how the instrument was to be used. Consignees were provided the option to exchange the unit for another model. The firm expanded the action on September 22, 2017, to request the return of the product. On September 27, 2017, the firm sent a notice to three distributors who were not identified in the original group of consignees. For further questions, please call (512) 381-7070.
Quantity
120 units
Distribution
US Distribution to the states of : FL, TX, NY, IL, GA, AL, MA, MO
Lot, serial or model codes
Lot numbers: 10923, 109231, 11476, 11993, WT20150630A
510(k) numbers
["K133245"]
PMA numbers
not on file
Product code
LRZ
Firm FEI
3008455034
Firm city
Bldg # 3 West Lake Hills
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register