Skip to the content

FDA Medical Device Recalls (openFDA)

CORNEAT VISION, LTD.: CorNeat EverPatch, REF: 1036295, Permanent Tissue-Integrating Surgical Matrix, 5.0 mm x 605 mm x 100 m

Recall number
Z-0039-2026
Recalling firm
CORNEAT VISION, LTD.
Product
CorNeat EverPatch, REF: 1036295, Permanent Tissue-Integrating Surgical Matrix, 5.0 mm x 605 mm x 100 m
Status
Open, Classified
Initiated
2025-08-18
Posted
2025-10-03
Terminated
not on file
Reason
Reviewed complaints have identified that conjunctival wound dehiscence may result in early exposure of the surgical patch.
Root cause
Unknown/Undetermined by firm
Action
On August 18, 2025, CorNeat Vision, Ltd. issued a "Urgent: Medical Device Recall" Notification to affected consignees via E-Mail. CorNeat ask consignees to take the following actions: 1. Review inventory, Quarantine and Discontinue use of All Lots of the CorNeat EverPatch and/or EverPatch+. 2. Please discard any unused product. 3. Eyecare providers should continue to monitor patients who have had CorNeat EverPatch and/or EverPatch+ implanted at the follow-up frequency necessary to detect signs of wound dehiscence or conjunctival thinning/erosion. 4. Patients should be instructed to seek prompt follow-up if they experience any signs or symptoms consistent with this complication. 5. This notice needs to be passed on to all those within your organization or to any organization where the product has been transferred. 6. Please fill in the MEDICAL DEVICE RECALL RETURN RESPONSE form at www.corneat.com/everpatch-recall or fill in the attached form and forward to CorNeat Vision by email to complaints@corneat.com .
Quantity
630 units
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CA, CO, CT, DC, DE, FL, GA, ID, IL, IN, KS, MA, MD, MI, MN, MO, MT, NC, NY, OH, PA, SC, TN, TX, UT, VA, WA, WI and the countries of Japan, Israel, Argentina, Hong Kong.
Lot, serial or model codes
All Lots/ UDI: G16010362950
510(k) numbers
["K223074"]
PMA numbers
not on file
Product code
QWU
Firm FEI
3029772936
Firm city
Ra'Anana
Firm state
not on file
Firm country
Israel

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register