Richard-Allan Scientific Company: Thermo Scientific Microm: Rotary Microtome: REF HM355S-3 Used for slicing tissue blocks.

Recall number
Z-0039-2012
Recalling firm
Richard-Allan Scientific Company
Product
Thermo Scientific Microm: Rotary Microtome: REF HM355S-3 Used for slicing tissue blocks.
Status
Terminated
Initiated
2011-08-12
Posted
2011-10-12
Terminated
2013-07-25
Reason
The firm has received two complaints of finger lacerations in operators of this device.
Root cause
Device Design
Action
ThermoFisher Scientific sent "URGENT MEDICAL DEVICE CORRECTION, RE: Thermo Scientific HM355S-3 User Safety Advisory Notice" to all affected customers. The letter states the manufacturer identified a need to enhance the safety instructions and the description of the safety devices in the user manual. It mentions the HM3555S-3 User Manual has been modified to include more specific safe work practices and a new Quick -Reference Card summarizing the manual's safety enhancements and that the changes have been included on a CD that is enclosed with the letter. The firm requests confirmation of receipt of the new User Manual and Quick-Reference Card by October 31, 2011 to be sent by FAX to (269) 372-2674 For questions on this recall call the firms Technical Support representatives at 1-800-522-7270 08:00 AM- 4:30 PM.
Quantity
786
Distribution
Worldwide Distribution: Nationwide distribution; including the country of Canada.
Lot, serial or model codes
Model HM355S-3, Serial Nos: 34450, 34519, 34520, 34601, 34662, 34665, 34807, 34922, 35030, 35032, 35033, 35159, 35637, 35793, 35976, 35978, 35978, 38337, 38584, 38651, 39238, 39280, 39281, 39542, 39795, 40607, 41039, 45104, 45367, 45369, 45371, 45760, 45845, 47466, 47467, 47468, 47469, 47772, 47773, 47774, 47775, 47776, 48095, 48096, 48097, 48098, 48109, 48110, 48111, 48112, 48113, 48114, 48115, 48116, 48184, 48252, 48253, 48254, 48259, 48260, 48261, 48314, 48315, 48593, 48594, 48910, 48911, 48912, 48913, 48914, 48915, 48916, 48917, 48918, 48919, 48920, 48921, 48922, 48923, 48924, 48978, 48978, 49102, 49103, 49104, 49105, 49106, 49107, 49108, 49109, 49110, 49209, 49209, 49210, 49210, 49211, 49211, 49212, 49213, 49213, 49214, 49214, 49221, 49223, 49230, 49286, 49287, 49288, 49289, 49290, 49291, 49292, 49293, 49294, 49295, 49296, 49297, 49298, 49299, 49300, 49301, 49302, 49303, 49304, 49305, 49306, 49307, 49308, 49309, 49310, 49311, 49312, 49962, 49963, 49964, 49965, 49966, 49967, 49968, 49969, 49970, 49971, 49972, 49973, 49974, 49975, 49976, 49977, 49978, 49989, 49990, 49991, 49992, 49993, 49994, 49995, 50221, 50222, 50224, 50225, 50226, 50578, 50579, 50580, 50581, 50582, 50583, 50584, 50585, 50586, 50587, 50676, 50677, 50678, 50679, 50680, 50681, 50791, 50792, 50793, 50794 and 50911.
510(k) numbers
not on file
PMA numbers
not on file
Product code
IDO
Firm FEI
1831638
Firm city
Kalamazoo
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28