Siemens Healthcare Diagnostics, Inc.: Dimension(R) EXL LOCI Cardiac Troponin I Calibrator (RC621) The LOCI Cardiac TNI Calibrator is an in vitro diagnostic product for the calibration on Cardiac Troponin I method on the Dimension(R) EXL (TM) Integrated chemistry system with LOCI(R) module.

Recall number
Z-0038-2012
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product
Dimension(R) EXL LOCI Cardiac Troponin I Calibrator (RC621) The LOCI Cardiac TNI Calibrator is an in vitro diagnostic product for the calibration on Cardiac Troponin I method on the Dimension(R) EXL (TM) Integrated chemistry system with LOCI(R) module.
Status
Terminated
Initiated
2011-08-31
Posted
2011-10-12
Terminated
2013-05-20
Reason
Internal testing has confirmed that recovery of levels 2 and 3 of LOCl Cardiac Troponin I Calibrator is lower than expected bottle value. Following calibration with LOCl Cardiac Troponin I Calibrator, lot 1AD045 an upward shift in patient and QC results may be observed.
Root cause
Other
Action
Siemens sent an Urgent Field Safety Notice dated August 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately discontinue use and discard any remaining inventory. Replacement orders would be processed and sent free of charge. Customers were also instructed to complete the attached FIELD CORRECTION EFFECTIVENESS CHECK form and fax it to (302) 631- 8457. Customers were instructed to forward the Safety Notice to anyone to whom they may have distributed the affected product. For any technical questions customers were to contact the Siemens Solutions Center at 800-441-9250.
Quantity
998
Distribution
Worldwide Distribution - USA (nationwide) including AL, AK, AZ, AR, CA, CO, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NM, NY, NC, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, and WY and the countries of GUAM, Austria, Canada, China, Czech Republic, Egypt, France, Germany, Italy, Poland, Portugal, Puerto Rico, Spain and Switzerland.
Lot, serial or model codes
Lot Number 1AD045
510(k) numbers
["K081683"]
PMA numbers
not on file
Product code
JJE
Firm FEI
2517506
Firm city
Newark
Firm state
DE
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register