Skip to the content

FDA Medical Device Recalls (openFDA)

Airgas Puritan Medical: AGI-2238 Portable 4 Cylinder Manifold Cart, for connecting medical gas cylinders.

Recall number
Z-0038-2010
Recalling firm
Airgas Puritan Medical
Product
AGI-2238 Portable 4 Cylinder Manifold Cart, for connecting medical gas cylinders.
Status
Terminated
Initiated
2009-08-06
Posted
2009-10-14
Terminated
2010-01-26
Reason
Yoke may become displaced. Yoke index safety system could become compromised, allowing a user to attach the wrong gas type to the yoke connection.
Root cause
Component design/selection
Action
Airgas issued a Medical Device Removal letter dated 8/6/09 to its customers informing them of the problem and the need to have a yoke back-up plate installed on the Manifold Block Assembly.
Quantity
208
Distribution
Nationwide distribution.
Lot, serial or model codes
All units of model numbers 8065-870 and 8066-910
510(k) numbers
not on file
PMA numbers
not on file
Product code
CAN
Firm FEI
not on file
Firm city
Radnor
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register