Skip to the content

FDA Medical Device Recalls (openFDA)

Canon Medical System, USA, INC.: This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.

Recall number
Z-0037-2026
Recalling firm
Canon Medical System, USA, INC.
Product
This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.
Status
Open, Classified
Initiated
2025-09-02
Posted
2025-10-08
Terminated
not on file
Reason
It has been found that when SPOT Fluoro is first turned on, the default ROI is selected and the collimator blades move to the boundary of the ROI, and the static image is overlayed around the ROI, covering the collimated area. If the customer changes the ROI, the collimator blades and static images are adjusted accordingly. However, when the acquisition program is changed, the collimator returns to the default ROI position which may be larger than the specified one, leaving the X-ray area partially covered by the static image. This is a software problem with the SPOT Fluoro program and will not occur if the function is not used.
Root cause
Radiation Control for Health and Safety Act
Action
Letters prepared and emailed to notify the customer of the defect. The letter contains a return reply form that is to be completed by the customer and emailed to CMSU for retention.
Quantity
8
Distribution
U.S.
Lot, serial or model codes
Serial Numbers: SZA2422001, SXA2432001, SUC23Z2014, BBF2392003, SUC2392013, SZA2422002, SXA2432002, SZA2442004.
510(k) numbers
["K210900","K233107"]
PMA numbers
not on file
Product code
OWB
Firm FEI
2020563
Firm city
Tustin
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register