QUANTUM SURGICAL SAS: The EPIONE device v1.0.2 is a user controlled, stereotactic accessory device intended to assist in the planning and manual advancement of one or more instruments, as well as in verification of instrument position during Computed Tomography (CT) guided percutaneous procedures. During the planning phase, the desired instrument placement and performance is defined relative to the target anatomy. During the guidance phase, the device enables to monitor respiratory levels and verify patient position prior to instrument advancement. During the assessment phase, the achieved instrument placement and performance are displayed relative to the previously defined plan through an overlay of the pre- and post-treatment image data.

Recall number
Z-0037-2025
Recalling firm
QUANTUM SURGICAL SAS
Product
The EPIONE device v1.0.2 is a user controlled, stereotactic accessory device intended to assist in the planning and manual advancement of one or more instruments, as well as in verification of instrument position during Computed Tomography (CT) guided percutaneous procedures. During the planning phase, the desired instrument placement and performance is defined relative to the target anatomy. During the guidance phase, the device enables to monitor respiratory levels and verify patient position prior to instrument advancement. During the assessment phase, the achieved instrument placement and performance are displayed relative to the previously defined plan through an overlay of the pre- and post-treatment image data.
Status
Completed
Initiated
2024-08-05
Posted
2024-10-09
Terminated
not on file
Reason
Quantum Surgical has become aware that a component (central axis) inside the Needle Guide product (product number 02-0027) is the subject of rust.
Root cause
Under Investigation by firm
Action
Quantum Surgical issued an URGENT FIELD SAFETY NOTICE to its sole consignee via visit on 08/05/2024 and followed up with an email on 08/19/2024. The notice explained the problem, potential risk, and requested that the consignee cease use of the device. The affected units were returned, repaired, and replaced.
Quantity
2 units
Distribution
US Nationwide distribution in the state of FL.
Lot, serial or model codes
UDI/DI 37603054002NGXL, serial numbers: 0321015 and 0321017
510(k) numbers
["K223758"]
PMA numbers
not on file
Product code
JAK
Firm FEI
3023140882
Firm city
montpellier
Firm state
not on file
Firm country
France

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register