FDA Medical Device Recalls (openFDA)
Xoran Technologies, Inc.: Xoran Mini CT Scanner
- Recall number
- Z-0036-2014
- Recalling firm
- Xoran Technologies, Inc.
- Product
- Xoran Mini CT Scanner
- Status
- Terminated
- Initiated
- 2010-07-27
- Posted
- not on file
- Terminated
- 2014-09-11
- Reason
- A description of the defect in the product or the manner in which the product fails to comply with an applicable Federal standard; The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xoran's specifications. See ATTACHMENT 1 for details. (e) A n evalu
- Root cause
- Radiation Control for Health and Safety Act
- Action
- All of the MiniCAT CT Scanners subject to these complaints were serviced by Xoran technicians and the out-of-specifications devices were brought into specification. This was accomplished by either readjusting the settings or by installing a new component.
- Quantity
- 320
- Distribution
- Worldwide Distribution - US nationwide
- Lot, serial or model codes
- Minicat
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- JAK
- Firm FEI
- 3004198450
- Firm city
- Ann Arbor
- Firm state
- MI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28