FDA Medical Device Recalls (openFDA)
LeMaitre Vascular, Inc.: Artegraft Collagen Vascular Grafts Model Numbers: (1) AG 616, (2) AG 630, (3) AG 636, (4) AG 640, (5) AG 645, (6) AG 715, (7) AG 730, (8) AG 735, (9) AG 740, (10) AG 745, (11) AG 750, (12) AG 830, (13) AG 840, (14) AG 1015, (15) AG 1030;
- Recall number
- Z-0034-2026
- Recalling firm
- LeMaitre Vascular, Inc.
- Product
- Artegraft Collagen Vascular Grafts Model Numbers: (1) AG 616, (2) AG 630, (3) AG 636, (4) AG 640, (5) AG 645, (6) AG 715, (7) AG 730, (8) AG 735, (9) AG 740, (10) AG 745, (11) AG 750, (12) AG 830, (13) AG 840, (14) AG 1015, (15) AG 1030;
- Status
- Open, Classified
- Initiated
- 2025-09-02
- Posted
- 2025-10-02
- Terminated
- not on file
- Reason
- The bovine carotid arteries used to produce the impacted grafts from affected lots were sourced using raw material from a supplier that had not been reviewed and approved by the appropriate regulatory authority.
- Root cause
- Nonconforming Material/Component
- Action
- On September 2, 2025 URGENT: MEDICAL DEVICE RECALL letters were sent to customers. Actions to be taken: 1. Check your inventory against the list of lots in this letter. Immediately quarantine any recalled devices. 2. Complete and return attached response form. 3. If the product has been implanted, graft explantation/patient intervention is NOT required. 4. When a recalled device has been returned to LeMaitre Vascular, a replacement device will be provided. 5. If you have transferred devices to another facility, please forward a copy of this recall letter to them. If you have any questions, contact Monte Nelson; mnelson@lemaitre.com; 732-422-8333 Monday through Friday, 8:00 AM to 4:30 PM, Eastern Time.
- Quantity
- 28 units
- Distribution
- US Nationwide distribution in the states of CA, CT, FL, GA, IL, MA, MS, NC, ND, NJ, NY, OK, PA, TN, TX, VA, WA.
- Lot, serial or model codes
- Model Numbers: (1) AG 616, (2) AG 630, (3) AG 636, (4) AG 640, (5) AG 645, (6) AG 715, (7) AG 730, (8) AG 735, (9) AG 740, (10) AG 745, (11) AG 750, (12) AG 830, (13) AG 840, (14) AG 1015, (15) AG 1030; UDI-DIs: (1) 00316837000213, (2) 00316837000220, (3) 00316837000237, (4) 00316837000244, (5) 00316837000251, (6) 00316837000268, (7) 00316837000275, (8) 00316837000282, (9) 00316837000299, (10) 00316837000305, (11) 00316837000312, (12) 00316837000367, (13) 00316837000329, (14) 00316837000343, (15) 00316837000350; Serial Numbers: (1) 23DD203-019, 23DD203-046, (2) 23DD203-024, 23DD203-029, (3) 23DD203-045, (4) 23DD203-018, (5) 23DD203-004, (6) 23DD203-021, 23DD203-044, 23DD203-038, 23DD203-039, 23DD203-011, 23DD203-032, 23DD203-031, 23DD203-043, (7) 23DD203-040, 23DD203-030, 23DD203-037, (8) 23DD203-035, 23DD203-033, (9) 23DD203-034, (10) 23DD203-036, (11) 23DD203-022, 23DD203-023, (12) 23DD203-017, (13) 23DD203-027, (14) 23KK498-014, (15) 23DD203-026;
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- LXA
- Firm FEI
- 1000121696
- Firm city
- North Brunswick
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28