Skip to the content

FDA Medical Device Recalls (openFDA)

Becton Dickinson & Company: BD Vacutainer Push Button Blood Collection Set with Pre-attached Holder. Non-pyrogenic 100. Rx only, sterile REF 368656 23G x 3/4'' x 12'' 0.6 x 19mm x 305mm ----- REF 367352 21G x 3/4'' x 12'' 0.8 x 19mm x 305mm

Recall number
Z-0034-2007
Recalling firm
Becton Dickinson & Company
Product
BD Vacutainer Push Button Blood Collection Set with Pre-attached Holder. Non-pyrogenic 100. Rx only, sterile REF 368656 23G x 3/4'' x 12'' 0.6 x 19mm x 305mm ----- REF 367352 21G x 3/4'' x 12'' 0.8 x 19mm x 305mm
Status
Terminated
Initiated
2006-08-11
Posted
2006-10-12
Terminated
2007-05-08
Reason
A complaint was received regarding a BD blood collection set where the blister was noticeably warped. The seal width was below minimum specification. Some blisters could adhere together and crack upon separation. Some units may have pre-activated or have defects affecting package integrity.
Root cause
Other
Action
Recall notification were mailed out first class mail, return receipt requested, approximately 8/24/2006 to distributors, who were then asked to contact their customers.
Quantity
110,300 units
Distribution
Nationwide.
Lot, serial or model codes
Catalog number 367352 Lot # 6100596 Catalog number 367352 Lot # 6100597 Catalog number 367352 Lot # 6152413 Catalog number 368656 Lot # 6090251 Catalog number 368656 Lot # 6094813 Catalog number 368656 Lot # 6094840 Catalog number 368656 Lot # 6111757 Catalog number 368656 Lot # 6121663 Catalog number 368656 Lot # 6146186
510(k) numbers
["K011984"]
PMA numbers
not on file
Product code
JAK
Firm FEI
2243072
Firm city
Franklin Lakes
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register