Skip to the content

FDA Medical Device Recalls (openFDA)

VIASYS Med Systems: Neonatal/Infant Hyperinflation Circuit, consisting of a 0.5 L breathing bag, pressure monitoring tube, oxygen supply tubing, sideport connector with adjustable bleed valve, ported elbow and infant mask; Mfg. for Corpak MedSystems, Wheeling, IL 60090; 20 units per case; reorder #BLD-83100-24

Recall number
Z-0034-04
Recalling firm
VIASYS Med Systems
Product
Neonatal/Infant Hyperinflation Circuit, consisting of a 0.5 L breathing bag, pressure monitoring tube, oxygen supply tubing, sideport connector with adjustable bleed valve, ported elbow and infant mask; Mfg. for Corpak MedSystems, Wheeling, IL 60090; 20 units per case; reorder #BLD-83100-24
Status
Terminated
Initiated
2003-09-10
Posted
2003-10-15
Terminated
2004-03-24
Reason
The exhalation hole in the sideport relief valve may be blocked, allowing excess pressure build-up.
Root cause
Other
Action
VIASYS telephoned the direct accounts and sent out follow-up faxes on 9/10-11/03, informing them of the potential for a blocked exhalation hole in the sideport relief valve cap which may result in excess pressure build-up. The accounts were instructed to check their inventory for the affected catalog numbers and lots numbers, place any found on hold for return to VIASYS, and sub-recall the affected product from their customers.
Quantity
2,720 units
Distribution
Missouri, Texas, New York, Illinois, Kansas, California, Pennsylvania, Georgia, Arizona, Massachusetts, Maine, Oklahoma and Virginia.
Lot, serial or model codes
reorder #BLD-83100-24, lots 11394 and 11680
510(k) numbers
["K930252"]
PMA numbers
not on file
Product code
CAI
Firm FEI
3011270181
Firm city
Wheeling
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register