Skip to the content

FDA Medical Device Recalls (openFDA)

Sentinel CH SpA: Alinity c CRP Vario Cardiac High Sensitivity Calibrator Kit (ref 07P5605) - Product Usage: use in the calibration of the Alinity c CRP Vario assay, Cardiac CRP high sensitivity method [Cardiac CRP], on the Alinity c analyzer.

Recall number
Z-0033-2022
Recalling firm
Sentinel CH SpA
Product
Alinity c CRP Vario Cardiac High Sensitivity Calibrator Kit (ref 07P5605) - Product Usage: use in the calibration of the Alinity c CRP Vario assay, Cardiac CRP high sensitivity method [Cardiac CRP], on the Alinity c analyzer.
Status
Terminated
Initiated
2021-07-29
Posted
not on file
Terminated
2023-07-24
Reason
Calibrator vials were incorrectly labeled. Use of incorrectly labeled vials will cause the instrument to generate a warning that calibrators are incorrectly loaded and calibration will not be completed successfully.
Root cause
Under Investigation by firm
Action
Firm notified distributor July 29, 2021, and provided an Urgent Field Safety Notice letter to be forwarded to final end users on August 20, 2021. Customers were advised to do the following: - Immediately discontinue use of affected lot of calibrator - Destroy any remaining inventory of affected lot according to your laboratory procedures - Contact Abbott to order replacement material - Review the content of the communication with your Medical Director - Retain the letter for future reference Please forward notification to all individuals in your organization who need to be aware, or to any individuals or organizations where the recalled lot was transferred. If you have questions regarding this recall, please contact your local area Customer Service.
Quantity
32
Distribution
US Nationwide distribution in the states of LA, MD, MN, ND, NY, WV, and Puerto Rico.
Lot, serial or model codes
Lot 00681Y600
510(k) numbers
not on file
PMA numbers
not on file
Product code
JIS
Firm FEI
3000303338
Firm city
Milan
Firm state
not on file
Firm country
Italy

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register