St Jude Medical Inc.: Merlin@Home Software Model EX2000 v8.2.2 for Merlin@Home Transmitter Models EX1150, EX1150W, EX1100, and EX1100W. The Merlin@home Transmitters are a component of St. Jude Medicals (SJM) Remote Care System. These Merlin@home transmitters are used to retrieve information from the SJM implantable medical device (IMD) using RF telemetry. The data is then sent back to the Merlin.net Patient Care Network through a built-in modem or, alternatively, through a USB cellular adapter or a USB based Broadband Kit. The transmitted data enables clinicians to remotely monitor a patients IMD using a secure, encrypted, password-protected web site, thus reducing the number of in-clinic visits a patient needs to make to their physician. The transmitter monitors the patients device daily between scheduled follow-ups and can alert a clinician if it detects anything out of the norm.

Recall number
Z-0033-2018
Recalling firm
St Jude Medical Inc.
Product
Merlin@Home Software Model EX2000 v8.2.2 for Merlin@Home Transmitter Models EX1150, EX1150W, EX1100, and EX1100W. The Merlin@home Transmitters are a component of St. Jude Medicals (SJM) Remote Care System. These Merlin@home transmitters are used to retrieve information from the SJM implantable medical device (IMD) using RF telemetry. The data is then sent back to the Merlin.net Patient Care Network through a built-in modem or, alternatively, through a USB cellular adapter or a USB based Broadband Kit. The transmitted data enables clinicians to remotely monitor a patients IMD using a secure, encrypted, password-protected web site, thus reducing the number of in-clinic visits a patient needs to make to their physician. The transmitter monitors the patients device daily between scheduled follow-ups and can alert a clinician if it detects anything out of the norm.
Status
Terminated
Initiated
2017-08-28
Posted
not on file
Terminated
2020-07-08
Reason
New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications.
Root cause
Unknown/Undetermined by firm
Action
Abbott sent an Important Cybersecurity Advisory dated August 28, 2017, to all affected customers to notify customers of the availability of the programmer software update and associated pacemaker firmware update. The notification material is in the form of a Physician Letter and Hospital Letter and will be delivered to physicians and hospitals with inventory in the US by overnight service. Customers with questions were instructed to contact their Abbott representative or the customers technical support hotline at 1-800-722-3774.
Quantity
495,011 units (391,093 in US)
Distribution
Worldwide Distribution
Lot, serial or model codes
Model EX1100 - GTIN 05414734504782 and Model EX1150 - GTIN 05414734404334
510(k) numbers
not on file
PMA numbers
["P910023S386"]
Product code
OSR
Firm FEI
2017865
Firm city
Sylmar
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register