Quality Electrodynamics LLC: TxRx 15CH Knee Coil 1.5T QED part number:Q7000016 Siemens Model # 10185453

Recall number
Z-0032-2020
Recalling firm
Quality Electrodynamics LLC
Product
TxRx 15CH Knee Coil 1.5T QED part number:Q7000016 Siemens Model # 10185453
Status
Terminated
Initiated
2019-08-20
Posted
not on file
Terminated
2020-05-26
Reason
Unintended ferrous material may be present within the pushbutton assembly of Siemens Knee and Siemens Extremity Coils.
Root cause
Nonconforming Material/Component
Action
The firm sent a notification letter to their customers who that have received the affected device. Customers were instructed to. 1. Immediately examine your inventory and quarantine product subject to recall. 2. if you have further distributed this product to MRI end users, please identify your affected customers and notify them. 3. Please ensure they understand to stop use of the coil for scanning and to notify you immediately.
Quantity
3 units
Distribution
Worldwide distribution - US in the states: NC OUS: Germany
Lot, serial or model codes
Serial # 1471 2106 1991
510(k) numbers
["K082636"]
PMA numbers
not on file
Product code
MOS
Firm FEI
3007350713
Firm city
Mayfield Village
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register