Oridion Medical 1987 Ltd.: Capnostream20 (US) N REFURBISHED, Part Number CS78651

Recall number
Z-0032-2019
Recalling firm
Oridion Medical 1987 Ltd.
Product
Capnostream20 (US) N REFURBISHED, Part Number CS78651
Status
Terminated
Initiated
2018-07-31
Posted
not on file
Terminated
2022-09-20
Reason
The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory default settings when the monitor is powered off.
Root cause
Nonconforming Material/Component
Action
Urgent Medical Device Correction notification letters dated 7/31/18 were distributed to customers. The letters included the following instructions for customers: Share this notification with all care environments where the Capnostream" 20 and Capnostream" 20p bedside patient monitors are used, particularly in the areas where nurse call may be enabled (general care floors). If your facility has distributed these bedside patient monitors to other persons or facilities, please promptly forward a copy of this letter to those recipients. Medtronic is recommending continued use of Capnostream" devices. Users should confirm that the date/time is accurately displayed. If the date/time requires reset, all user defined settings, such as alarms and nurse call, will also require reset until the software is updated. Please follow the instruction in the Operators Manual to set user-defined default settings. The Operators Manual is available at http://www.medtronic.com/covidien/enus/support/product-manuals.html, refer to the Institutional Settings section. Return the completed Acknowledgement and Receipt Form by fax or email even if you have no inventory. In October 2018, download the software update from the Medtronic website via the following link which includes directions on how to download and install the software: http://www.medtronic.com/covidien/en-us/support/patient-monitoring-equipmentsoftware-upgrades.html.
Quantity
46
Distribution
The products were distributed US nationwide and Puerto Rico. The products were distributed to the following foreign countries: Argentina, Australia, Austria, Bahrain, Belgium, Canada, Columbia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Kazakhstan, Kuwait, Mexico, Netherlands, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, UAE, UK, and Vietnam.
Lot, serial or model codes
B200001005 B200001300 B200001477 B200001546 B200001548 B200001557 B200001560 B200001582 B200001583 B200001591 B200001598 B200001601 B200001603 B200001606 B200001631 B200001641 B200001817 B200001884 B200001920 B200001938 B200001977 B200001986 B200002051 B200002060 B200002070 B200002078 B200002134 B200002148 B200002163 B200001387 B200001569 B200001673 B200001679 B200001844 B200002025 B200002031 B200002145 B200001827 B200001137 B200001160 B200001310 B200001660 B200001872 B200002187 B400002132 B200001671
510(k) numbers
not on file
PMA numbers
not on file
Product code
DQA
Firm FEI
3001227743
Firm city
Jerusalem
Firm state
Har Hotzvim Industrial Park
Firm country
Israel

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register